background image

10 - ENG    PN MN236 A | User Manual

Sprint / Stroller / Hi Flow Stroller

NOTE: If the unit is empty, only the red colored band 
of the scale will be exposed. If this is the case, the 
unit must be filled prior to use.

7. Under certain environmental conditions, with 

continuous use, and especially with the portable 

not being moved, it may develop excessive frost 

around the vaporizer coil and on the case. You 

may reduce this frost by tapping the unit and/or 

wiping any accumulated frost off the case.

Condensate Pad  
Maintenance

After the unit is empty and has warmed up to 

room temperature, remove any moisture from the 

condensation cup:

For hard case

• Tilt the unit to one side. Grasp the knob at the 

center of the condensation cup (at the bottom 

of the unit) and pull straight out. This will 

release the cup.

NOTE: If heavily soiled, the pad can be machine 
washed (with bleach to disinfect).

• To replace the cup, center it over the round 

opening at the bottom of the unit and press 

the knob in one click to engage the cup and 

then a second click to lock the cup in place.

     

    Knob Cup       

NOTE: If the knob will not push in and engage the 
cup, hold the cup and pull the knob out until it clicks 
into the open position. Now replace the cup as previ-
ously described.

For soft case

• Unzip the bottom zipper and pull back the 

bottom cover.

• Remove the pad from the cup and wring out 

any absorbed moisture. Allow the pad to dry 

completely before replacing it in the cup.

NOTE: If heavily soiled, the pad can be machine 
washed (with bleach to disinfect).

• To replace, simply insert pad into bottom 

cover and rezip the bottom zipper.

NOTE: If the knob will not push in and engage the 
cup, hold the cup and pull the knob out until it clicks 
into the open position. Now replace the cup as previ-
ously described.

Frost Reduction

Under certain environmental conditions, with 

continuous use, and especially with the portable 

not being moved, it may develop excessive frost 

around the vaporizer coil and on the case. You may 

reduce this frost by tapping the unit and/or wiping 

any accumulated frost off the case.

WARNING: DUE TO THE POSSIBILITY OF 
EXCESSIVE MOISTURE / ICE BUILD UP IT 
IS RECOMMENDED THAT THE HI FLOW 
STROLLER REMAIN IDLE FOR 1 HOUR 
AFTER USE TO DECREASE MOISTURE 
BUILDUP.

Due to the higher flow rates associated with the Hi 

Flow Stroller, condensation and some frosting will 

occur on the unit, particularly on flow settings of 

10 LPM and higher. It is recommended that the unit 

be allowed at least 1 hour of inactivity after use to 

decrease the possibility of excessive build up due to 

continuous usage.

Summary of Contents for SPRINT

Page 1: ...Sprint Stroller Hi Flow Stroller User Manual English...

Page 2: ...Warning Reg W001 Council Directive 93 42 EEC concerning medical devices Authorized representative in the European Community This device complies with the requirements of Directive 93 42 EEC concerning...

Page 3: ...pacity 513 L 1026 L 1025 L Weight Empty 4 5 lb 2 04 kg 5 0 lb 2 27 kg 5 5 lb 2 49 kg Weight Filled 6 0 lb 2 72 kg 8 0 lb 3 63 kg 8 5 lb 3 86 kg Height 11 75 in 298 mm 13 5 in 343 mm 13 5 in 343 mm Dim...

Page 4: ...on the order of a physician The unit contains liquid oxygen which is extremely cold almost 300 F 184 C Exposure to such a low temperature can cause severe frostbite Liquid and gaseous oxygen though no...

Page 5: ...capabilities It is recommended to have an alternate source of supplemental oxygen in the event of mechani cal failure These portable devices also have a quick disconnect valve that can be coupled with...

Page 6: ...off 0 position 3 Position the fill connector in the upright position over the stationary unit s connector 4 a If your connector is a push on Lower the Sprint Stroller Hi Flow Stroller unit on to the...

Page 7: ...al NOTE Liquid level gauge readings are most accurate 30 seconds or more after the Sprint Stroller Hi Flow Stroller is filled 7 You may also see vapor around the connection This is normal and is due t...

Page 8: ...the Sprint Stroller Hi Flow Stroller is being refilled immediately after a period of use fill it and then wait 10 minutes after filling with the FCV set at zero for the pressure to stabilize Then oper...

Page 9: ...oxygen in the unit with the electronic liquid level gauge Gen 3 Gauge Depress the push button liquid level switch on top of the unit for two seconds minimum Read across the top of the light bar to in...

Page 10: ...sition Now replace the cup as previ ously described For soft case Unzip the bottom zipper and pull back the bottom cover Remove the pad from the cup and wring out any absorbed moisture Allow the pad t...

Page 11: ...e off 0 position The Portable does not fill Verify that there is oxygen in the Reservoir Ensure that the Portable and Reservoir fill connectors are fully engaged through out the filling process The Po...

Page 12: ...l Always return Sprint Stroller Hi Flow Stroller in cluding all components to your homecare provider for proper disposal You can also contact your local city or town offices for instructions on proper...

Page 13: ...gnetic compatibility EMC and needs to be installed and put into service according to the EMC information provided in this manual Caution Portable and mobile radio frequency RF communications equipment...

Page 14: ...standard requirement for equipment which has been tested to specific test levels and over specific frequency ranges and been found compliant with regulations Table 3 Guidance and manufacturers declara...

Page 15: ...om fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equip ment marked...

Page 16: ...h no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 30107 U...

Reviews: