8 - ENG    PN MN236 A | User Manual

Sprint / Stroller / Hi Flow Stroller

Operating Instructions

(Cannula is not included.)

1. Push the cannula breathing tube firmly on to the 

oxygen tube connector.

2. Adjust your cannula breathing tube to the proper 

position so that you will be able to breathe the 

oxygen comfortably.

3. Turn the flow control knob on top of the Sprint/

Stroller/Hi Flow Stroller/Hi Flow Stroller clock-

wise until the prescribed flow rate (numeral) 

is visible in the knob “window” and a positive 

detent is felt.

Caution: Out-of-specification oxygen 
flow will result if the flow control knob 
is set between flow rates.

4. You should be receiving oxygen now. To stop 

flow of oxygen, turn the flow control knob coun-

terclockwise to the off (0) position.

Caution:  Always turn the flow control 
knob off (0) when not in use.

Caution: Check liquid level gauge only 
after the valve is closed.

Caution: If the Sprint/Stroller/Hi Flow Stroller is 
being refilled immediately after a period of use, 
fill it, and then wait 10 minutes after filling with 
the FCV set at zero for the pressure to stabilize. 
Then operate as normal.

WARNING: DO NOT OPERATE SPRINT/
STROLLER/HI FLOW STROLLER WHILE AT-
TACHED TO THE STATIONARY UNIT.

WARNING: IF THE SPRINT/STROLLER/HI FLOW 
STROLLER DOES NOT SEPARATE EASILY, DO NOT 
USE FORCE. THE UNITS MAY BE FROZEN TOGETH-
ER. LEAVE THE UNITS CONNECTED AND WAIT 
UNTIL THE UNITS WARM UP — THEN THEY WILL 
SEPARATE EASILY. DO NOT TOUCH ANY FROSTED 
PARTS. WARNING: SHOULD THE LEAKAGE BE 
EXCESSIVE TO THE POINT THAT A STREAM OF 
LIQUID IS PRESENT, LEAVE THE AREA AND CALL 
YOUR HEALTH CARE PROVIDER IMMEDIATELY.
WARNING: LIQUID OR COLD GASEOUS OXYGEN 
CAN FROSTBITE THE SKIN. OXYGEN WILL RESULT 
IN INCREASED FIRE HAZARD.

Caution: Should there be any liquid leak-
age from the stationary unit or Sprint/
Stroller/Hi Flow Stroller separating the 
units, set the Sprint/Stroller/Hi Flow 
Stroller aside, ensuring it remains verti-
cal, leave the room, and call your health 
care provider immediately.

Summary of Contents for SPRINT

Page 1: ...Sprint Stroller Hi Flow Stroller User Manual English...

Page 2: ...Warning Reg W001 Council Directive 93 42 EEC concerning medical devices Authorized representative in the European Community This device complies with the requirements of Directive 93 42 EEC concerning...

Page 3: ...pacity 513 L 1026 L 1025 L Weight Empty 4 5 lb 2 04 kg 5 0 lb 2 27 kg 5 5 lb 2 49 kg Weight Filled 6 0 lb 2 72 kg 8 0 lb 3 63 kg 8 5 lb 3 86 kg Height 11 75 in 298 mm 13 5 in 343 mm 13 5 in 343 mm Dim...

Page 4: ...on the order of a physician The unit contains liquid oxygen which is extremely cold almost 300 F 184 C Exposure to such a low temperature can cause severe frostbite Liquid and gaseous oxygen though no...

Page 5: ...capabilities It is recommended to have an alternate source of supplemental oxygen in the event of mechani cal failure These portable devices also have a quick disconnect valve that can be coupled with...

Page 6: ...off 0 position 3 Position the fill connector in the upright position over the stationary unit s connector 4 a If your connector is a push on Lower the Sprint Stroller Hi Flow Stroller unit on to the...

Page 7: ...al NOTE Liquid level gauge readings are most accurate 30 seconds or more after the Sprint Stroller Hi Flow Stroller is filled 7 You may also see vapor around the connection This is normal and is due t...

Page 8: ...the Sprint Stroller Hi Flow Stroller is being refilled immediately after a period of use fill it and then wait 10 minutes after filling with the FCV set at zero for the pressure to stabilize Then oper...

Page 9: ...oxygen in the unit with the electronic liquid level gauge Gen 3 Gauge Depress the push button liquid level switch on top of the unit for two seconds minimum Read across the top of the light bar to in...

Page 10: ...sition Now replace the cup as previ ously described For soft case Unzip the bottom zipper and pull back the bottom cover Remove the pad from the cup and wring out any absorbed moisture Allow the pad t...

Page 11: ...e off 0 position The Portable does not fill Verify that there is oxygen in the Reservoir Ensure that the Portable and Reservoir fill connectors are fully engaged through out the filling process The Po...

Page 12: ...l Always return Sprint Stroller Hi Flow Stroller in cluding all components to your homecare provider for proper disposal You can also contact your local city or town offices for instructions on proper...

Page 13: ...gnetic compatibility EMC and needs to be installed and put into service according to the EMC information provided in this manual Caution Portable and mobile radio frequency RF communications equipment...

Page 14: ...standard requirement for equipment which has been tested to specific test levels and over specific frequency ranges and been found compliant with regulations Table 3 Guidance and manufacturers declara...

Page 15: ...om fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equip ment marked...

Page 16: ...h no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 30107 U...

Reviews: