4 - ENG    PN MN236 A | User Manual

Sprint / Stroller / Hi Flow Stroller

Warning Information

Important:

 Read this manual thoroughly before 

operating the Sprint/Stroller/Hi Flow Stroller.

RX Only.

WARNING: THIS DEVICE IS NOT IN-
TENDED FOR LIFE SUSTAINING USE.

WARNING: IF YOU FEEL THE EQUIPMENT IS NOT 
OPERATING PROPERLY, CALL YOUR HEALTH CARE 
PROVIDER. DO NOT ATTEMPT TO REPAIR OR 
ADJUST THE UNIT YOURSELF.
WARNING: DO NOT MODIFY THIS EQUIPMENT 
WITHOUT AUTHORIZATION FROM THE MANUFAC-
TURER.
WARNING: IF CONTINUITY OF OXYGEN SUPPLY IS 
REQUIRED, ENSURE THAT AN ADEQUATE SUPPLY 
OF OXYGEN AND/OR A SECONDARY OXYGEN 
SUPPLY IS AVAILABLE AT ALL TIMES DURING 
THERAPY.

WARNING: DO NOT ALLOW SMOKING, 
CANDLES, OR OPEN FLAMES WITHIN 10 
FEET (3 METERS) OF THE DEVICE, OR 
CLOSER THAN 8 INCHES (20 CM) FROM 
ANY SOURCE OF IGNITION.
WARNING: KEEP YOUR UNIT IN A WELL-
VENTILATED AREA.
WARNING: DO NOT STORE LIQUID 
OXYGEN EQUIPMENT IN A CLOSET, CAR 
TRUNK, OR OTHER CONFINED AREA. DO 
NOT PLACE BLANKETS, DRAPERIES, OR 
OTHER FABRICS OVER EQUIPMENT.

WARNING: THIS PRODUCT CAN EXPOSE YOU 
TO CHEMICALS INCLUDING NICKEL, WHICH IS 
KNOWN TO THE STATE OF CALIFORNIA TO CAUSE 
CANCER.  FOR MORE INFORMATION, GO TO 
WWW.P65WARNINGS.CA.GOV.

Caution: Use the Sprint/Stroller/Hi Flow 
Stroller only as directed by your doctor.

Caution: Federal law restricts this device to sale 
by or on the order of a physician.

The unit contains liquid oxygen which is 

extremely cold, almost -300°F (-184°C). 

Exposure to such a low temperature can 

cause severe frostbite.
Liquid and gaseous oxygen, though 

nonflammable, cause other materials to 

burn faster than normal. This hazard, 

along with the low temperature of liquid 

oxygen, warrants certain safety precau-

tions.

Keep flammable materials away from 

this equipment. Aerosol sprays, oils and 

grease, including facial creams and pe-

troleum jelly, ignite easily and may burn 

rapidly in the presence of oxygen.
Smoking while wearing an oxygen can-

nula can cause facial burns and possibly 

result in death.

Removing the cannula and placing it on 

clothing, bedding, sofas, or other cush-

ion material will cause a flash fire when 

exposed to a cigarette, heat source, 

spark or flame. 

If you smoke please: (1) turn off the 

portable, (2) take off the cannula, and 

(3) leave the room where the device is 

located.
In the event of an accidental tip-over, 

immediately but cautiously return the 

unit into an upright position if possible. 

If any liquid oxygen is escaping, leave 

the area immediately and call your 

healthcare provider. Do not attempt 

to move the unit or stop the liquid 

oxygen from escaping.

Do not touch frosted parts of any unit.

Do not store or operate the portable 

coupled to the stationary unit.
Do not allow untrained personnel to 

handle or operate this device.
Use of this device is prohibited on com-

mercial passenger and cargo air flights 

by the Federal Aviation Administration.

Summary of Contents for SPRINT

Page 1: ...Sprint Stroller Hi Flow Stroller User Manual English...

Page 2: ...Warning Reg W001 Council Directive 93 42 EEC concerning medical devices Authorized representative in the European Community This device complies with the requirements of Directive 93 42 EEC concerning...

Page 3: ...pacity 513 L 1026 L 1025 L Weight Empty 4 5 lb 2 04 kg 5 0 lb 2 27 kg 5 5 lb 2 49 kg Weight Filled 6 0 lb 2 72 kg 8 0 lb 3 63 kg 8 5 lb 3 86 kg Height 11 75 in 298 mm 13 5 in 343 mm 13 5 in 343 mm Dim...

Page 4: ...on the order of a physician The unit contains liquid oxygen which is extremely cold almost 300 F 184 C Exposure to such a low temperature can cause severe frostbite Liquid and gaseous oxygen though no...

Page 5: ...capabilities It is recommended to have an alternate source of supplemental oxygen in the event of mechani cal failure These portable devices also have a quick disconnect valve that can be coupled with...

Page 6: ...off 0 position 3 Position the fill connector in the upright position over the stationary unit s connector 4 a If your connector is a push on Lower the Sprint Stroller Hi Flow Stroller unit on to the...

Page 7: ...al NOTE Liquid level gauge readings are most accurate 30 seconds or more after the Sprint Stroller Hi Flow Stroller is filled 7 You may also see vapor around the connection This is normal and is due t...

Page 8: ...the Sprint Stroller Hi Flow Stroller is being refilled immediately after a period of use fill it and then wait 10 minutes after filling with the FCV set at zero for the pressure to stabilize Then oper...

Page 9: ...oxygen in the unit with the electronic liquid level gauge Gen 3 Gauge Depress the push button liquid level switch on top of the unit for two seconds minimum Read across the top of the light bar to in...

Page 10: ...sition Now replace the cup as previ ously described For soft case Unzip the bottom zipper and pull back the bottom cover Remove the pad from the cup and wring out any absorbed moisture Allow the pad t...

Page 11: ...e off 0 position The Portable does not fill Verify that there is oxygen in the Reservoir Ensure that the Portable and Reservoir fill connectors are fully engaged through out the filling process The Po...

Page 12: ...l Always return Sprint Stroller Hi Flow Stroller in cluding all components to your homecare provider for proper disposal You can also contact your local city or town offices for instructions on proper...

Page 13: ...gnetic compatibility EMC and needs to be installed and put into service according to the EMC information provided in this manual Caution Portable and mobile radio frequency RF communications equipment...

Page 14: ...standard requirement for equipment which has been tested to specific test levels and over specific frequency ranges and been found compliant with regulations Table 3 Guidance and manufacturers declara...

Page 15: ...om fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equip ment marked...

Page 16: ...h no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 30107 U...

Reviews: