HELiOS Portables Technical Service Manual • PN 20562190 Rev E
22
HELiOS P
or
tables
Maintenance
X
Schedule B – Continuous (As Needed) Schedule
A. Introduction
Continuous maintenance is a set of tests and inspections to be
done periodically to ensure that the equipment is functioning
properly. It can be done while the equipment is in service by
drivers or other personnel.
1. If a unit fails a given test, it should be taken out of service
and sent to the repair center/department for further inspec-
tion.
2. Schedule – The continuous schedule should be performed
every time the driver visits a patient or when their reser-
voir is filled or replaced. It should also be performed when
equipment is transferred between patients.
B. Procedure
1. Ensure the unit is empty of liquid oxygen contents.
2. Visual Inspection
a. Inspect the case, flow knob, and vent valve lever for
cracks, warping, or discoloration.
b. Verify that the label on the rear cover is present and leg-
ible.
c. Verify that the fill connector poppet is not worn or bro-
ken.
d. Verify that the lip seal in the fill connector is not cracked
or showing signs of wear.
e. Move the vent valve lever down to the open position.
Verify that it moves smoothly when opening and clos-
ing the vent valve.
3. Clean and disinfect the unit if necessary using the procedures
described in the Operations section of this manual.
4. Fill the H300 or H850 with LOX.
5. Post-Fill Inspection
a. Verify that the QDV poppet is closed and not leaking.
b. Verify that the Vent Valve is closed and not leaking.
c. Carefully pull up on the contents indicator strap.
Verify that it moves smoothly and reads accurately.
d. Flow Control Valve Test (RP19)