background image

eQuinox User Manual  —  11

eQuinox

12VDC Cable

12VDC Cable

A 12VDC Cable allows the system to operate from DC outlets, 

such as those found in motor vehicles. 
  1. Start your vehicle. 
  2. Insert the 12 Volt DC cord into the recessed power 

receptacle on the side of the device. 

  3. Insert the 12VDC Cable plug into the DC power outlet in 

your motor vehicle.

When the device is properly connected and receiving power from 

the DC power source, a green indicator light on the Power Supply 

and the External Power Indicator             on the control panel will 

illuminate.

SeQual eQuinox DC Supply Connection

When the eQuinox is connected to the 12VDC Cable in a vehicle, 

the following applies:

 The eQuinox will run at 3.0LPM on continuous flow while 

being operated with the DC Supply in a car.

 The eQuinox will charge the battery while on 12VDC Cable in 

a car while running on continuos flow at 2LPM or less.

 If the eQuinox battery is being charged, the battery symbol will 

waterfall. If the eQuinox battery is not charging, the battery 

symbol will be displayed solid.

Settings Available When Operating on DC for eQuinox

Flow Setting

Continuous

Bolus Size (mL)

Pulse

0.5

Yes

-

-

1.0

Yes

16

Yes

1.5

Yes

24

Yes

2.0

Yes

32

Yes

2.5

Yes

40

Yes

3.0

Yes

48

Yes

-

-

56

Yes

-

-

64

Yes

-

-

72

Yes

-

-

80

Yes

-

-

88

Yes

-

-

96

Yes

-

-

128

Yes

-

-

160

Yes

-

-

192

Yes

Note: While operating the device from a 12VDC Cable, the Battery may 
not charge.
Note: When operating the eQuinox from a 12VDC Cable, it is recommend-
ed that the Pulse Dose Mode be used if prescribed to use a pulse dose 
device.
Note: To ensure that the device is utilizing the DC power, check that the 
External Power Indicator is illuminated. If a beep is heard, this could indi-
cate a loss of external power. Check if the green Battery Power Verification 
LED is on or off. If the battery is removed on DC power, it will ensure no 
loss of charge while the unit is running. If the indicator is not illuminated, 
the battery is powering the device.

WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE 
AIR OUTLET VENTS TO BECOME BLOCKED. THIS CAN CAUSE 
THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR 

PERFORMANCE.
WARNING: IF THE VEHICLE’S DC POWER SOURCE DROPS BELOW 11.5 
VOLTS, THE EQUINOX WILL REVERT TO THE BATTERY OPERATION (IF 
THE BATTERY IS PRESENT AND CHARGED). (THE EXTERNAL POWER 
INDICATOR MAY REMAIN ON.)

Note: Actual performance of the eQuinox may vary depending upon usage 
conditions of the motor vehicle.

Summary of Contents for eQuinox

Page 1: ...eQuinoxTM User Manual eQuinox...

Page 2: ...Output Do not get wet Type B Applied Part degree of protection against electric shock the applicable U S and Canadian standards Recycle Symbol This device complies with the requirements of Directive 9...

Page 3: ...ations for Use 7 Contraindications for Use 7 Safety Guidelines 7 Recommended Operating Environments 9 Operating Instructions 10 First Time eQuinox Operation 14 User Troubleshooting Table 16 Cleaning C...

Page 4: ...s Cautions and additional device information in the rest of this manual Become familiar with the key features of the eQuinox and the User Control Panel BACK Power Cartridge Latch Exhaust Vent Power Ca...

Page 5: ...Pulse Dose Mode Indicator illuminates and a pulse of oxygen is delivered with each inspiratory effort ALERT Yellow Indicator Low and Medium Priority Alerts When illuminated this indicates a low priori...

Page 6: ...g the eQuinox 7 Universal Cart Operation Flow Mode Button and Indicator Pressing this button repeatedly will toggle you back and forth between 0 5 and 3 0 LPM In Pulse Dose Mode oxygen is supplied in...

Page 7: ...S WITHIN THE SAME ROOM OF 1 THIS DEVICE OR 2 ANY OXYGEN CARRYING ACCESSORY FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE PROPERTY DAMAGE AND OR CAUSE PHYSICAL INJURY OR DEATH WARNING DO NO...

Page 8: ...XYGEN CONCENTRATOR IN THE PRESENCE OF FLAMMABLE GASES THIS CAN RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE BODILY INJURIES OR DEATH WARNING REMOVING THE CANNULA AND PUTTING IT ON CLOTH ING BEDDING...

Page 9: ...rmal operating temperature before being turned on Refer to the Specifications section in this manual Note When the automobile in which you are using the Oxygen Con centrator unit is turned off disconn...

Page 10: ...wer source and replenish the battery Proper connection of the AC Supply or 12VDC cable requires aligning the power plug with the recessed external power receptacle on the eQuinox The recessed external...

Page 11: ...Flow Setting Continuous Bolus Size mL Pulse 0 5 Yes 1 0 Yes 16 Yes 1 5 Yes 24 Yes 2 0 Yes 32 Yes 2 5 Yes 40 Yes 3 0 Yes 48 Yes 56 Yes 64 Yes 72 Yes 80 Yes 88 Yes 96 Yes 128 Yes 160 Yes 192 Yes Note W...

Page 12: ...if not fully charged or too warm If the Battery is not present or fully discharged during an AC power interruption the eQuinox will shut down and alarm The Battery is fully discharged when the Power...

Page 13: ...ry charges when in the device is operating on AC power only The Battery will also charge from the optional Desktop Charger Typical New Power Cartridge Operating Times Operating Type Battery Type 12 ce...

Page 14: ...ulse Dose Mode and Continuous Flow Mode The proper mode and proper setting should be prescribed for you by your physician Continuous Flow Mode Operation When operating in the Continuous Flow Mode a co...

Page 15: ...for use with the Oxygen Concentrator Nasal Cannula with 7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www...

Page 16: ...re adequate air flow Filter restricted Clean or replace Air Inlet Filter Position your eQuinox to ensure adequate air flow Exceeding maximum breath rate in Pulse Dose mode See Pulse Dose mode Breath R...

Page 17: ...you with cleaning disinfection and replacement information Cleaning Care Routine Maintenance CAUTION Do not operate the Oxygen Concentrator with out the air intake filter in place If a second filter i...

Page 18: ...Cable Power Cord and Outer Case Turn OFF the eQuinox and disconnect from AC or DC power before any cleaning or disinfection activity DO NOT spray the outer case directly Use a damp not wet cloth or s...

Page 19: ...92 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse Dose Mode only Pulse Dose Mode Power Cartridge Airline Travel Table mL SETTING FLIGHT LENGTH in Hrs During taxi take off and la...

Page 20: ...82 2473 Worldwide www CAIREmedical com Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany Worldwide www CAIREmedical com PN 4805 C3_01 A Chart SeQual Technologies Inc reserves the righ...

Reviews: