eQuinox

User Troubleshooting Table 

Do Not Ignore Alarms

Symptom

Possible Cause

Your Action

eQuinox does not power 
on when ON/OFF button is 
pressed

No battery installed

Install battery or plug into external power 

Battery is discharged or no 
external power is present

Plug into external power

AC Power Supply or 12v 
DC cable is not plugged in 
correctly

Check connections at outlet, Power Supply box, and at the eQuinox to ensure secure 
connections.

Insufficient power from the 
outlet

Ensure there is adequate power from the outlet. (Outlet may be controlled by a wall 
switch or the circuit breaker may have engaged.)

Other

Contact your dealer/provider

No Oxygen

Filter blocked

Clean Air Inlet Filter

Humidifier

Check humidifier attachment and tubing

eQuinox not ON

Power the SeQual eQuinox ON

Tubing or cannula is kinked or 
not properly connected 

Check tubing, cannula, and connections

Other

Contact your dealer/provider

Low Oxygen 
Concentration

Restriction in tubing or 
humidifier

Repair or replace tubing or humidifier.  Position your eQuinox to ensure adequate air 
flow.

Filter restricted

Clean or replace Air Inlet Filter. Position your eQuinox to ensure adequate air flow.

Exceeding maximum breath 
rate in Pulse Dose mode. See 
Pulse Dose mode Breath Rate 
Chart

Switch to Continuous Flow mode until respiratory rate reaches an acceptable rate.

Other

Contact your dealer/provider

Low Oxygen Flow

Restriction in humidifier or 
tubing

Repair or replace humidifier or tubing

Filter Blocked

Clean or replace Air Inlet Filter.

Other

Contact your dealer/provider

No Oxygen Delivered in Pulse 
Flow Mode

Tubing/cannula longer than 7 
feet (2.1m)

Attach 7 foot (2.1m) tubing/cannula

Humidifier attached
No inspiration detected

Contact your dealer/provider

Unit in Continuous Flow mode Switch to Pulse Dose mode
Nasal cannula not in the nose, 
or mouth breathing

Place nasal cannula in nostrils and breathe through nose

Power Cartridge Status Gauge 
never indicates fully charged

Power Cartridge (battery) is 
aging

Contact your dealer/provider to replace Power cartridge (battery)

WARNING: THE MANUFACTURER RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT OF A POWER OUT-
AGE, ALARM CONDITION, OR MECHANICAL FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE 
SYSTEM REQUIRED.

Remove humidifier

Summary of Contents for eQuinox

Page 1: ...eQuinoxTM User Manual eQuinox...

Page 2: ...Output Do not get wet Type B Applied Part degree of protection against electric shock the applicable U S and Canadian standards Recycle Symbol This device complies with the requirements of Directive 9...

Page 3: ...ations for Use 7 Contraindications for Use 7 Safety Guidelines 7 Recommended Operating Environments 9 Operating Instructions 10 First Time eQuinox Operation 14 User Troubleshooting Table 16 Cleaning C...

Page 4: ...s Cautions and additional device information in the rest of this manual Become familiar with the key features of the eQuinox and the User Control Panel BACK Power Cartridge Latch Exhaust Vent Power Ca...

Page 5: ...Pulse Dose Mode Indicator illuminates and a pulse of oxygen is delivered with each inspiratory effort ALERT Yellow Indicator Low and Medium Priority Alerts When illuminated this indicates a low priori...

Page 6: ...g the eQuinox 7 Universal Cart Operation Flow Mode Button and Indicator Pressing this button repeatedly will toggle you back and forth between 0 5 and 3 0 LPM In Pulse Dose Mode oxygen is supplied in...

Page 7: ...S WITHIN THE SAME ROOM OF 1 THIS DEVICE OR 2 ANY OXYGEN CARRYING ACCESSORY FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE PROPERTY DAMAGE AND OR CAUSE PHYSICAL INJURY OR DEATH WARNING DO NO...

Page 8: ...XYGEN CONCENTRATOR IN THE PRESENCE OF FLAMMABLE GASES THIS CAN RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE BODILY INJURIES OR DEATH WARNING REMOVING THE CANNULA AND PUTTING IT ON CLOTH ING BEDDING...

Page 9: ...rmal operating temperature before being turned on Refer to the Specifications section in this manual Note When the automobile in which you are using the Oxygen Con centrator unit is turned off disconn...

Page 10: ...wer source and replenish the battery Proper connection of the AC Supply or 12VDC cable requires aligning the power plug with the recessed external power receptacle on the eQuinox The recessed external...

Page 11: ...Flow Setting Continuous Bolus Size mL Pulse 0 5 Yes 1 0 Yes 16 Yes 1 5 Yes 24 Yes 2 0 Yes 32 Yes 2 5 Yes 40 Yes 3 0 Yes 48 Yes 56 Yes 64 Yes 72 Yes 80 Yes 88 Yes 96 Yes 128 Yes 160 Yes 192 Yes Note W...

Page 12: ...if not fully charged or too warm If the Battery is not present or fully discharged during an AC power interruption the eQuinox will shut down and alarm The Battery is fully discharged when the Power...

Page 13: ...ry charges when in the device is operating on AC power only The Battery will also charge from the optional Desktop Charger Typical New Power Cartridge Operating Times Operating Type Battery Type 12 ce...

Page 14: ...ulse Dose Mode and Continuous Flow Mode The proper mode and proper setting should be prescribed for you by your physician Continuous Flow Mode Operation When operating in the Continuous Flow Mode a co...

Page 15: ...for use with the Oxygen Concentrator Nasal Cannula with 7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www...

Page 16: ...re adequate air flow Filter restricted Clean or replace Air Inlet Filter Position your eQuinox to ensure adequate air flow Exceeding maximum breath rate in Pulse Dose mode See Pulse Dose mode Breath R...

Page 17: ...you with cleaning disinfection and replacement information Cleaning Care Routine Maintenance CAUTION Do not operate the Oxygen Concentrator with out the air intake filter in place If a second filter i...

Page 18: ...Cable Power Cord and Outer Case Turn OFF the eQuinox and disconnect from AC or DC power before any cleaning or disinfection activity DO NOT spray the outer case directly Use a damp not wet cloth or s...

Page 19: ...92 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse Dose Mode only Pulse Dose Mode Power Cartridge Airline Travel Table mL SETTING FLIGHT LENGTH in Hrs During taxi take off and la...

Page 20: ...82 2473 Worldwide www CAIREmedical com Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany Worldwide www CAIREmedical com PN 4805 C3_01 A Chart SeQual Technologies Inc reserves the righ...

Reviews: