background image

14  —  4805-C3_01 A

eQuinox

Step 1: Positioning Your eQuinox for Use 

Place the eQuinox in a well-ventilated, well lit area. Be sure the 

air inlet and exhaust vents are not obstructed.

Position the eQuinox so that all 

audible and visual indicators or 

alarms can be easily seen and 

heard.

place before operating your 

missing, contact your home care 

dirty, wash with warm soapy 

water, rinse with clear water and 

allow it to dry before placing the 

If necessary, replace it with a new, 

Plug the device into a grounded 

AC Power outlet, a DC Power source, or be sure there is a fully 

charged Battery installed. 

Step 2:  Power ON the Device and Allow it to 
Warm-Up

Press and hold the “ON/OFF” Button for one (1) second to power 

ON your eQuinox. A brief visual power-on self-test will occur. 

Step 3:  Connect the Oxygen Supply Tubing or Nasal 
Cannula to the Oxygen Outlet

Replace the cannula and oxygen supply tubing regularly, as recommended by 
your home care provider. Consult your home care provider for cannula and 
supply tubing cleaning and replacement information.

Note:  Ensure the cannula is fully inserted and secure. This ensures that the 
Oxygen Concentrator can properly detect inspiration for oxygen delivery. 
During inhalation, you should hear or feel oxygen flow to the prongs of 
the nasal cannula.

Step 4: 

 

Select the Flow Mode

Press the Flow Mode button to select the desired mode. Pressing 

this button repeatedly will toggle you back and forth between 

Pulse Dose Mode and Continuous Flow Mode. The proper 

mode and proper setting should be prescribed for you by your 

physician.

Continuous Flow Mode Operation:

When operating in the Continuous Flow Mode, a continuous 

when in Continuous Flow Portable Concentrator Mode.

Pulse Dose Mode Operation:

When operating in Pulse Mode, a bolus of oxygen, measured 

in milliliters (mL), is delivered upon inspiration. Remove the 

the Pulse Dose Mode. DO NOT lay the eQuinox down while 

To ensure proper saturations, secure a prescription for pulse dose 

from your physician with consideration to your needs at rest, 

during exercise, and when traveling at altitude. Contact your care 

provider for prescription and boarding forms.

When operating in pulse dose mode, the eQuinox supplies a 

will continually check for an inspiratory effort every 15 seconds. 

Once an inspiration is detected, the device reverts to measured 

bolus delivery. To exit Pulse Dose Mode and engage Continuous 

First Time eQuinox Operation

The eQuinox has an oxygen concentration status indicator 

(OCSI) built in to the device. The OCSI continually monitors 

the oxygen output of the device and incorporates a green light 

(normal), yellow light (alert), red light (alarm) and a buzzer. 

These indicators will all illuminate during the power-on self-test 

at start-up. After initially powering ON the device, please allow 

Note:  If there are three beeps at start-up, this indicates a low battery 
condition for the 9-volt battery inside the cabinet of the eQuinox. Please 
contact your home care provider and request that the 9-volt battery be 
replaced. 

Air inlet Filter

Exhaust Vent

eQuinox. If the air inlet filter is

Summary of Contents for eQuinox

Page 1: ...eQuinoxTM User Manual eQuinox...

Page 2: ...Output Do not get wet Type B Applied Part degree of protection against electric shock the applicable U S and Canadian standards Recycle Symbol This device complies with the requirements of Directive 9...

Page 3: ...ations for Use 7 Contraindications for Use 7 Safety Guidelines 7 Recommended Operating Environments 9 Operating Instructions 10 First Time eQuinox Operation 14 User Troubleshooting Table 16 Cleaning C...

Page 4: ...s Cautions and additional device information in the rest of this manual Become familiar with the key features of the eQuinox and the User Control Panel BACK Power Cartridge Latch Exhaust Vent Power Ca...

Page 5: ...Pulse Dose Mode Indicator illuminates and a pulse of oxygen is delivered with each inspiratory effort ALERT Yellow Indicator Low and Medium Priority Alerts When illuminated this indicates a low priori...

Page 6: ...g the eQuinox 7 Universal Cart Operation Flow Mode Button and Indicator Pressing this button repeatedly will toggle you back and forth between 0 5 and 3 0 LPM In Pulse Dose Mode oxygen is supplied in...

Page 7: ...S WITHIN THE SAME ROOM OF 1 THIS DEVICE OR 2 ANY OXYGEN CARRYING ACCESSORY FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE PROPERTY DAMAGE AND OR CAUSE PHYSICAL INJURY OR DEATH WARNING DO NO...

Page 8: ...XYGEN CONCENTRATOR IN THE PRESENCE OF FLAMMABLE GASES THIS CAN RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE BODILY INJURIES OR DEATH WARNING REMOVING THE CANNULA AND PUTTING IT ON CLOTH ING BEDDING...

Page 9: ...rmal operating temperature before being turned on Refer to the Specifications section in this manual Note When the automobile in which you are using the Oxygen Con centrator unit is turned off disconn...

Page 10: ...wer source and replenish the battery Proper connection of the AC Supply or 12VDC cable requires aligning the power plug with the recessed external power receptacle on the eQuinox The recessed external...

Page 11: ...Flow Setting Continuous Bolus Size mL Pulse 0 5 Yes 1 0 Yes 16 Yes 1 5 Yes 24 Yes 2 0 Yes 32 Yes 2 5 Yes 40 Yes 3 0 Yes 48 Yes 56 Yes 64 Yes 72 Yes 80 Yes 88 Yes 96 Yes 128 Yes 160 Yes 192 Yes Note W...

Page 12: ...if not fully charged or too warm If the Battery is not present or fully discharged during an AC power interruption the eQuinox will shut down and alarm The Battery is fully discharged when the Power...

Page 13: ...ry charges when in the device is operating on AC power only The Battery will also charge from the optional Desktop Charger Typical New Power Cartridge Operating Times Operating Type Battery Type 12 ce...

Page 14: ...ulse Dose Mode and Continuous Flow Mode The proper mode and proper setting should be prescribed for you by your physician Continuous Flow Mode Operation When operating in the Continuous Flow Mode a co...

Page 15: ...for use with the Oxygen Concentrator Nasal Cannula with 7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www...

Page 16: ...re adequate air flow Filter restricted Clean or replace Air Inlet Filter Position your eQuinox to ensure adequate air flow Exceeding maximum breath rate in Pulse Dose mode See Pulse Dose mode Breath R...

Page 17: ...you with cleaning disinfection and replacement information Cleaning Care Routine Maintenance CAUTION Do not operate the Oxygen Concentrator with out the air intake filter in place If a second filter i...

Page 18: ...Cable Power Cord and Outer Case Turn OFF the eQuinox and disconnect from AC or DC power before any cleaning or disinfection activity DO NOT spray the outer case directly Use a damp not wet cloth or s...

Page 19: ...92 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse Dose Mode only Pulse Dose Mode Power Cartridge Airline Travel Table mL SETTING FLIGHT LENGTH in Hrs During taxi take off and la...

Page 20: ...82 2473 Worldwide www CAIREmedical com Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany Worldwide www CAIREmedical com PN 4805 C3_01 A Chart SeQual Technologies Inc reserves the righ...

Reviews: