background image

eQuinox User Manual  —  17

eQuinox

Air Inlet Filter

Ambient air is drawn into the device through the air inlet located 

most important maintenance activity you will perform as a user.

Cleaning the air inlet filter will keep your eQuinox performing

properly. Check this air inlet filter daily and clean it at least once

a week using the following procedure: 
  1. Remove the air inlet filter 
  2. Wash the filter in warm tap water using a mild soap 

detergent solution

  3. Rinse the filter thoroughly with clear tap water and squeeze 

out the excess water

  4. Allow the filter to air dry
  5. Reinsert the filter in the cabinet

Routine Maintenance

Servicing of the internal components inside the cabinet of the

eQuinox must be conducted by a CAIRE trained and qualified

service technician as needed. 

Preventative Maintenance (PM) to be completed on an annual 

or yearly basis. Consult your provider to arrange PM schedule. 

inspections of the alarms, internal 9-volt battery, battery, internal 

User Care and Cleaning of the Device

Cannula Replacement

Replace your supply tubing and cannula on a regular basis as 

recommended by your home care provider. Your physician or 

home care provider will provide you with cleaning, disinfection 

and replacement information.

Cleaning, Care & Routine Maintenance

CAUTION: Do not operate the Oxygen Concentrator with-
out the air intake filter in place. If a second filter is pro-
vided, insert the “replacement” filter before you clean the 

dirty filter. Clean the dirty filter in a warm soap and water solution 
then dry thoroughly prior to use.

Note: The Manufacturer does not recommend the sterilization of this 
equipment.
Note: Do not operate the Oxygen Concentrator without the air intake
filter in place. If a second filter is provided, insert the “replacement” filter 
before you clean the dirty filter. Clean the dirty filter in a warm soap and 
water solution then dry thoroughly prior to use.

Note: Always follow the cannula manufacturer’s instructions for proper 
use. Replace the disposable cannula as recommended by the cannula 
manufacturer or your Equipment Provider. Additional supplies are avail-
able from your Equipment Provider.

Note: If three beeps sound after the power ON self-test, this indicates the 
9v battery is low in power. Contact your home care provider if you hear 
three beeps.

Air inlet

Filter

The air inlet filter should be replace annually. If the eQuinox is

used in a dusty environment, the filter may need to be replaced

more frequently. You should not operate the eQuinox without an

air inlet filter installed for more than 5 minutes.

Summary of Contents for eQuinox

Page 1: ...eQuinoxTM User Manual eQuinox...

Page 2: ...Output Do not get wet Type B Applied Part degree of protection against electric shock the applicable U S and Canadian standards Recycle Symbol This device complies with the requirements of Directive 9...

Page 3: ...ations for Use 7 Contraindications for Use 7 Safety Guidelines 7 Recommended Operating Environments 9 Operating Instructions 10 First Time eQuinox Operation 14 User Troubleshooting Table 16 Cleaning C...

Page 4: ...s Cautions and additional device information in the rest of this manual Become familiar with the key features of the eQuinox and the User Control Panel BACK Power Cartridge Latch Exhaust Vent Power Ca...

Page 5: ...Pulse Dose Mode Indicator illuminates and a pulse of oxygen is delivered with each inspiratory effort ALERT Yellow Indicator Low and Medium Priority Alerts When illuminated this indicates a low priori...

Page 6: ...g the eQuinox 7 Universal Cart Operation Flow Mode Button and Indicator Pressing this button repeatedly will toggle you back and forth between 0 5 and 3 0 LPM In Pulse Dose Mode oxygen is supplied in...

Page 7: ...S WITHIN THE SAME ROOM OF 1 THIS DEVICE OR 2 ANY OXYGEN CARRYING ACCESSORY FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE PROPERTY DAMAGE AND OR CAUSE PHYSICAL INJURY OR DEATH WARNING DO NO...

Page 8: ...XYGEN CONCENTRATOR IN THE PRESENCE OF FLAMMABLE GASES THIS CAN RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE BODILY INJURIES OR DEATH WARNING REMOVING THE CANNULA AND PUTTING IT ON CLOTH ING BEDDING...

Page 9: ...rmal operating temperature before being turned on Refer to the Specifications section in this manual Note When the automobile in which you are using the Oxygen Con centrator unit is turned off disconn...

Page 10: ...wer source and replenish the battery Proper connection of the AC Supply or 12VDC cable requires aligning the power plug with the recessed external power receptacle on the eQuinox The recessed external...

Page 11: ...Flow Setting Continuous Bolus Size mL Pulse 0 5 Yes 1 0 Yes 16 Yes 1 5 Yes 24 Yes 2 0 Yes 32 Yes 2 5 Yes 40 Yes 3 0 Yes 48 Yes 56 Yes 64 Yes 72 Yes 80 Yes 88 Yes 96 Yes 128 Yes 160 Yes 192 Yes Note W...

Page 12: ...if not fully charged or too warm If the Battery is not present or fully discharged during an AC power interruption the eQuinox will shut down and alarm The Battery is fully discharged when the Power...

Page 13: ...ry charges when in the device is operating on AC power only The Battery will also charge from the optional Desktop Charger Typical New Power Cartridge Operating Times Operating Type Battery Type 12 ce...

Page 14: ...ulse Dose Mode and Continuous Flow Mode The proper mode and proper setting should be prescribed for you by your physician Continuous Flow Mode Operation When operating in the Continuous Flow Mode a co...

Page 15: ...for use with the Oxygen Concentrator Nasal Cannula with 7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www...

Page 16: ...re adequate air flow Filter restricted Clean or replace Air Inlet Filter Position your eQuinox to ensure adequate air flow Exceeding maximum breath rate in Pulse Dose mode See Pulse Dose mode Breath R...

Page 17: ...you with cleaning disinfection and replacement information Cleaning Care Routine Maintenance CAUTION Do not operate the Oxygen Concentrator with out the air intake filter in place If a second filter i...

Page 18: ...Cable Power Cord and Outer Case Turn OFF the eQuinox and disconnect from AC or DC power before any cleaning or disinfection activity DO NOT spray the outer case directly Use a damp not wet cloth or s...

Page 19: ...92 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse Dose Mode only Pulse Dose Mode Power Cartridge Airline Travel Table mL SETTING FLIGHT LENGTH in Hrs During taxi take off and la...

Page 20: ...82 2473 Worldwide www CAIREmedical com Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany Worldwide www CAIREmedical com PN 4805 C3_01 A Chart SeQual Technologies Inc reserves the righ...

Reviews: