eQuinox User Manual  —  15

eQuinox

Pulse Dose Mode and autoSat Features

Your eQuinox has a feature called autoSAT

®

 that delivers a 

consistent-sized bolus of oxygen up to 40 breaths per minute (see 

Max Breath Rate table below). As your breath rate increases, ie. 

during ambulation, the autoSAT feature servo-controls the device, 

delivering the set bolus size. autoSAT Technology is proprietary 

to CAIRE and is utilized during Pulse Mode only.

Mas Breath Rate

Bolus 

V

olume (+/- 15%)

Pulse Dose Setting

Bolus Size

(± 15%) mL

Max Breath Rate

1.0

16

40

2.0

32

40

3.0

48

40

4.0

64

40

5.0

80

37

6.0

96

31

-

128

23

-

160

18

-

192

15

Step 5: 

 

Adjust the Flow Setting to the Prescribed 

Level

Using the Flow Setting Buttons, adjust the flow setting to the

prescribed setting supplied by your physician.

Note: Bolus volume decreases as breath rate exceeds published range.
Note: The pulse dose setting (bolus) is not equal to LPM.

WARNING: PULSE DOSE MODE SETTINGS MUST BE DETER-
MINED FOR EACH PATIENT INDIVIDUALLY FOR THEIR NEEDS

 

AT REST, DURING EXERCISE AND WHEN TRAVELING.

WARNING: THE EQUINOX MAY NOT BE ABLE TO DETECT ALL IN-
SPIRATORY EFFORTS IN PULSE MODE. IF THE EQUINOX DOES NOT 
RESPOND TO YOUR INSPIRATION EFFORTS, ASK YOUR HOME CARE 
PROVIDER TO CHECK THE SENSITIVITY SETTING.

Note: To Equipment Provider:  The following oxygen administration acces

-

sories are recommended for use with the Oxygen Concentrator: 

• Nasal Cannula with 7 feet (2.1 m) of tubing

: PN 5408-SEQ

• 

For any additional recommended accessories, please see the Accessories 

Catalog (PN ML-LOX0010) available on www.CAIREmedical.com.

WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE 
PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE 
FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO 

DO SO BY

 

A LICENSED CLINICIAN.

Step 6: 

 

Begin Using Your eQuinox

Breathe normally through the nasal cannula.

Note:  You will achieve longer operating time on the Battery if you operate 
your eQuinox in the Pulse Mode (Physician approved mode).

Step 7:  eQuinox Cart Operation

The eQuinox Cart is designed for use on pedestrian-type surfaces.

Align the locater holes on the bottom of the device to the tabs on 

the Universal Cart.

Align the large threaded screw to the 

insert at the back of the eQuinox. Then 

hand-tighten the knob on the cart until the 

Universal Cart is tightly secured to the 

SeQual eQuinox Concentrator.

Depress the push-button on the Universal 

Cart handle to adjust the height of the 

handle.

Step 8:  Power Off the Device

Press and hold the “ON/OFF” Button for two (2) seconds to 

power OFF your eQuinox.

Store the eQuinox in a cool and dry location for a period no 

longer than 1 month. The eQuinox must run for a minimum of 2 

hours per month to ensure proper function of the unit.

Note:  DO NOT operate the eQuinox and ambulate while a humidifier is 
attached. Remove the humidifier bottle before walking. DO NOT lay the 
eQuinox down while attached to a humidifier bottle.

Summary of Contents for eQuinox

Page 1: ...eQuinoxTM User Manual eQuinox...

Page 2: ...Output Do not get wet Type B Applied Part degree of protection against electric shock the applicable U S and Canadian standards Recycle Symbol This device complies with the requirements of Directive 9...

Page 3: ...ations for Use 7 Contraindications for Use 7 Safety Guidelines 7 Recommended Operating Environments 9 Operating Instructions 10 First Time eQuinox Operation 14 User Troubleshooting Table 16 Cleaning C...

Page 4: ...s Cautions and additional device information in the rest of this manual Become familiar with the key features of the eQuinox and the User Control Panel BACK Power Cartridge Latch Exhaust Vent Power Ca...

Page 5: ...Pulse Dose Mode Indicator illuminates and a pulse of oxygen is delivered with each inspiratory effort ALERT Yellow Indicator Low and Medium Priority Alerts When illuminated this indicates a low priori...

Page 6: ...g the eQuinox 7 Universal Cart Operation Flow Mode Button and Indicator Pressing this button repeatedly will toggle you back and forth between 0 5 and 3 0 LPM In Pulse Dose Mode oxygen is supplied in...

Page 7: ...S WITHIN THE SAME ROOM OF 1 THIS DEVICE OR 2 ANY OXYGEN CARRYING ACCESSORY FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE PROPERTY DAMAGE AND OR CAUSE PHYSICAL INJURY OR DEATH WARNING DO NO...

Page 8: ...XYGEN CONCENTRATOR IN THE PRESENCE OF FLAMMABLE GASES THIS CAN RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE BODILY INJURIES OR DEATH WARNING REMOVING THE CANNULA AND PUTTING IT ON CLOTH ING BEDDING...

Page 9: ...rmal operating temperature before being turned on Refer to the Specifications section in this manual Note When the automobile in which you are using the Oxygen Con centrator unit is turned off disconn...

Page 10: ...wer source and replenish the battery Proper connection of the AC Supply or 12VDC cable requires aligning the power plug with the recessed external power receptacle on the eQuinox The recessed external...

Page 11: ...Flow Setting Continuous Bolus Size mL Pulse 0 5 Yes 1 0 Yes 16 Yes 1 5 Yes 24 Yes 2 0 Yes 32 Yes 2 5 Yes 40 Yes 3 0 Yes 48 Yes 56 Yes 64 Yes 72 Yes 80 Yes 88 Yes 96 Yes 128 Yes 160 Yes 192 Yes Note W...

Page 12: ...if not fully charged or too warm If the Battery is not present or fully discharged during an AC power interruption the eQuinox will shut down and alarm The Battery is fully discharged when the Power...

Page 13: ...ry charges when in the device is operating on AC power only The Battery will also charge from the optional Desktop Charger Typical New Power Cartridge Operating Times Operating Type Battery Type 12 ce...

Page 14: ...ulse Dose Mode and Continuous Flow Mode The proper mode and proper setting should be prescribed for you by your physician Continuous Flow Mode Operation When operating in the Continuous Flow Mode a co...

Page 15: ...for use with the Oxygen Concentrator Nasal Cannula with 7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www...

Page 16: ...re adequate air flow Filter restricted Clean or replace Air Inlet Filter Position your eQuinox to ensure adequate air flow Exceeding maximum breath rate in Pulse Dose mode See Pulse Dose mode Breath R...

Page 17: ...you with cleaning disinfection and replacement information Cleaning Care Routine Maintenance CAUTION Do not operate the Oxygen Concentrator with out the air intake filter in place If a second filter i...

Page 18: ...Cable Power Cord and Outer Case Turn OFF the eQuinox and disconnect from AC or DC power before any cleaning or disinfection activity DO NOT spray the outer case directly Use a damp not wet cloth or s...

Page 19: ...92 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse Dose Mode only Pulse Dose Mode Power Cartridge Airline Travel Table mL SETTING FLIGHT LENGTH in Hrs During taxi take off and la...

Page 20: ...82 2473 Worldwide www CAIREmedical com Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany Worldwide www CAIREmedical com PN 4805 C3_01 A Chart SeQual Technologies Inc reserves the righ...

Reviews: