DT5730/DT5725 User Manual rev. 2
25
➢
DPP-PSD Control Software
1.
Launch DPP-PSD Control Software
2.
Connect to the digitizer
3.
Before to start the acquisition, go to the “Stats” tab and monitor the chann
el temperatures displayed in the
relevant column until you see they
don’t vary significantly
4.
Go to the “General” tab and press the “Calibrate” button
5.
Start the acquisition
Fig. 9.6:
Channel calibration in DPP-PSD Control Software
➢
MC
2
Analyzer
1.
Launch MC
2
A
2.
Connect to the digitizer
3.
Before to start the acquisition, monitor the channel temperatures in
Tools->ADC calibration
4.
Press “Calibrate” button to perform the calibration
5.
Start the acquisition
Fig. 9.7:
Channel calibration in MC
2
Analyzer software