Communication Protocol 

The following Protocol is based on commands made of ASCII characters strings.

 

Board commands 

The general board commands have the format: $CMD:***,PAR:***** <CR><LF> 

$CMD:SET,PAR:BDCLR 

Clears board alarms 

$CMD:MON,PAR:BDILK 

Monitor Interlock status 

$CMD:MON,PAR:BDNCH 

Number of channels 

$CMD:MON,PAR:BDNAME 

Board name 

$CMD:MON,PAR:BDSNUM 

Serial Number 

$CMD:MON,PAR:BDFREL 

Firmware release 

$CMD:MON,PAR:BDALARM 

Alarm Mask = 0x1ff8 

IS_OVC  

8  

At least one ch. in Overcurrent 

IS_OVV 

0x10  

At least one ch. in OverVoltage 

IS_UNV  

0x20  

At least one ch. in UnderVoltage 

IS_MAXV 

0x40 

At least one ch. in MaxV 

IS_TRIP 

0x80 

At least one ch. in Trip 

IS_MAXPW 

0x100 

At least one ch. in MaxPower 

IS_TWARN  

0x200 

At least one ch. in Temp warning (>80°C) 

IS_OVT 

0x400 

At least one ch. in OVT (>125°C) 

IS_KILL 

0x800 

At least one ch. Killed 

IS_INTLCK  

0x1000   At least one ch. in Interlock 

$CMD:MON,PAR:MACADDR 

Board MAC Address 

$CMD:MON,PAR:IPADDR 

Board IP Address 

$CMD:MON,PAR:SUBMASK 

Subnet mask 

$CMD:MON,PAR:GATEWAY 

Gateway 

$CMD:MON,PAR:DHCPEN 

DHCP (Enable / Disable) 

Channel Commands 

The channel commands have the format: $CMD:***,CH:*,PAR:***,VAL:xxx.xx<CR,LF> 
Where: 
CMD : SET, MON 

CH : Channel (0..4); CH = 4 “group commands”

 

PAR : VSET, ISET, ON ... 
VAL : value 

Response string has the format : #CMD:***,VAL:xxx.xx<CR,LF>  
Where: 
CMD : OK, ERR 
VAL : value 
In case of MON commands related to all channels, the returned value are separated by ';' 

Monitor commands 

$CMD:MON,CH:XX,PAR:VSET 

VSET value 

$CMD:MON,CH:XX,PAR:VSRES 

Resolution of VSET in Volt 

$CMD:MON,CH:XX,PAR:VSDEC 

Decimal digits of VSET 

$CMD:MON,CH:XX,PAR:VMAX 

Max value of VSET 

$CMD:MON,CH:XX,PAR:VMIN 

Min value of VSET 

$CMD:MON,CH:XX,PAR:VMON 

VMON value 

$CMD:MON,CH:XX,PAR:VMRES 

VMON resolution 

$CMD:MON,CH:XX,PAR:VMDEC 

Decimal digits of VMON 

$CMD:MON,CH:XX,PAR:ISET 

ISET value 

$CMD:MON,CH:XX,PAR:ISRES 

Resolution of ISET in µA 

$CMD:MON,CH:XX,PAR:IMAXH 

Max value of ISET in high range 

$CMD:MON,CH:XX,PAR:IMAXL 

Max value of ISET in low range 

Summary of Contents for DT5519EN

Page 1: ...CAEN Tools for Discovery n Electronic Instrumentation Es DT55xxE Desktop HV Power Supply Rev 6 11 April 2017...

Page 2: ...roduct is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsi...

Page 3: ...r 11 Power Supply Modules LabVIEW Instrument Driver 11 4 VT emulator control 12 Module access 12 Ethernet configuration 12 Firmware upgrade 13 Format EEPROM 14 Channels settings 14 Communication Proto...

Page 4: ...EP 6 kV 1 mA 4 W Positive HV outputs are delivered through SHV connectors Mixed versions have two positive and two negative channels The HV output RAMP UP and RAMP DOWN rates may be selected independe...

Page 5: ...cable 10BASE T Ethernet cable Power requirements The module is powered by the external AC DC stabilized power supply included in the delivered kit Switchbox Model FRA045 S12 4 see specifications in t...

Page 6: ...3 F GHz plug and jack Test voltage 10kV DC 1mn unmated connectors Ratings 12kV DC 1mn mated pairs Current rating 10 A WARNING These connectors produce extremely hazardous high voltages at a potential...

Page 7: ...trol panel NAME TYPE SIGNAL FUNCTION ALARM RED LED AMP 280371 2 Out Alarm status signalled active LOW INTERLOCK RED LED AMP 280371 2 In Interlock signal channels hardware disabled Alarm signal As an A...

Page 8: ...FUNCTION 12V In RAPC722X SWITCHCRAFT PCB 2 1mm DC Power Jack Input 12V DC Input ON OFF switch Molveno A11331122000 A1 switch O power supply OFF I power supply ON IMON Range The channel current can be...

Page 9: ...nA 2 of read 5nA 2 of read 500pA 2 of read 100nA 2 of read 50nA ISET vs IMON High range 2 of read 1 A 2 of read 50nA 2 of read 5nA 2 of read 1 A 2 of read 500nA Low range 2 of read 100nA 2 of read 5nA...

Page 10: ...arts DT55xxE desktop power supply unit External AC DC power supply with separate power cord USB cable 10BASE T Ethernet cable Moreover an external PC is required supported OS Windows 7 or later Linux...

Page 11: ...r level application such as GECO2020 and the lower layer software libraries For more info please visit www caen it products firmware software section Power Supply Modules LabVIEW Instrument Driver Pow...

Page 12: ...k and set the virtual communication port associated to the module Select USB connection and the used port number set port as follows baud rate 9600 Data bits 8 Parity none stop bit 1 Flow control none...

Page 13: ...in the EEPROM and go back to Main Menu the DHCP server will automatically assign a new IP to the module at next Power On Firmware upgrade To upgrade the firmware download from www caen it unit page th...

Page 14: ...ded firmware Format EEPROM By typing F on Terminal Board Menu Fig 10 it is possible to access the format EEPROM menu Fig 12 Format EEPROM Menu Channels settings By typing D on Terminal Board Menu Fig...

Page 15: ...0x1000 At least one ch in Interlock CMD MON PAR MACADDR Board MAC Address CMD MON PAR IPADDR Board IP Address CMD MON PAR SUBMASK Subnet mask CMD MON PAR GATEWAY Gateway CMD MON PAR DHCPEN DHCP Enable...

Page 16: ...hannel in Temperature warning 80 C IS_OVT 0x400 Channel in OVT 125 C IS_KILL 0x800 Channel Killed IS_INTLCK 0x1000 Channel in Interlock CMD MON CH XX PAR RUP RAMP UP value CMD MON CH XX PAR RUPMIN RAM...

Page 17: ...nge of customers including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Ve...

Reviews: