CAEN
Electronic Instrumentation
DT55xxE Desktop HV Power Supply
14
•
Select
“
File
”
> send file
•
Browse the image file (for example DT55XXE.102)
•
Select “open”
•
Wait the upload to complete until this message is shown:
•
Select
“
File
”
> disconnect
•
unplug the USB cable
•
connect once again to the module, via USB
•
now the unit is ready to operate running the upgraded firmware
Format EEPROM
By typing F on Terminal Board Menu (Fig. 10) it is possible to access the format EEPROM menu.
Fig. 12: Format EEPROM Menu
Channels settings
By typing D on Terminal Board Menu (Fig. 10) it is possible to access channels settings.
To
change one parameter: point the parameter with the “arrow keys”, and type the desired value, confirm
by pressing <Enter>; Imon range, Power and Power Down can be changed using the <Space> bar.
Fig. 13: Channels Menu
By typing G,
the channels are accessed in “group” mode (one setting applies to all channels).