background image

CAEN

 

 

Electronic Instrumentation

 

 

DT55xxE Desktop HV Power Supply

 

12 

4.

 

VT emulator control 

Module access 

The module is provided with a USB2.0 compliant interface and Ethernet Port (TCP/IP).  
For link connection, refer to instructions given at page 10.  
CAEN provides the CAEN GECO2020 Control Software that allows a friendly remote management of all 

Unit’s functional parameters (see 

www.caen.it software support page); anyway, the connection can be 

performed also via VT emulators, we suggest to use Tera Term (see https://ttssh2.osdn.jp/index.html.en) 

USB 

Launch the terminal emulator (we suggest using 
Tera Term), select the serial communication link 
and  set  the  virtual  communication  port 
associated to the module. Select USB connection 
and the used port number; set port as follows 

baud rate 9600 

Data bits: 8 

Parity: none 

stop bit: 1 

Flow control: none 

Launch communication 

Type caen then <enter>; 

Ethernet 

Launch  the  terminal  emulator  (we  suggest  using 
Tera Term), select port number 

23

 

Please note that line editing must be disabled prior 
to Ethernet access. 
Default settings are: 

IP address  

192.168.0.1  

Subnet mask  

255.255.255.0 

Gateway  

255.255.255.0 

Launch communication 
Type caen then <enter>; 

 
as the communication is established, the Main Menu will be displayed: 

  

Fig. 10: Terminal Board Menu 

Five options are available: 

 

Display 

 

Format 

 

Update 

 

Ethernet 

 

Quit 

Ethernet configuration 

To configure the Ethernet Port; with reference to  Fig. 10: Terminal Board Menu, Type E; the following 
screen will open: 

Summary of Contents for DT5519EN

Page 1: ...CAEN Tools for Discovery n Electronic Instrumentation Es DT55xxE Desktop HV Power Supply Rev 6 11 April 2017...

Page 2: ...roduct is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsi...

Page 3: ...r 11 Power Supply Modules LabVIEW Instrument Driver 11 4 VT emulator control 12 Module access 12 Ethernet configuration 12 Firmware upgrade 13 Format EEPROM 14 Channels settings 14 Communication Proto...

Page 4: ...EP 6 kV 1 mA 4 W Positive HV outputs are delivered through SHV connectors Mixed versions have two positive and two negative channels The HV output RAMP UP and RAMP DOWN rates may be selected independe...

Page 5: ...cable 10BASE T Ethernet cable Power requirements The module is powered by the external AC DC stabilized power supply included in the delivered kit Switchbox Model FRA045 S12 4 see specifications in t...

Page 6: ...3 F GHz plug and jack Test voltage 10kV DC 1mn unmated connectors Ratings 12kV DC 1mn mated pairs Current rating 10 A WARNING These connectors produce extremely hazardous high voltages at a potential...

Page 7: ...trol panel NAME TYPE SIGNAL FUNCTION ALARM RED LED AMP 280371 2 Out Alarm status signalled active LOW INTERLOCK RED LED AMP 280371 2 In Interlock signal channels hardware disabled Alarm signal As an A...

Page 8: ...FUNCTION 12V In RAPC722X SWITCHCRAFT PCB 2 1mm DC Power Jack Input 12V DC Input ON OFF switch Molveno A11331122000 A1 switch O power supply OFF I power supply ON IMON Range The channel current can be...

Page 9: ...nA 2 of read 5nA 2 of read 500pA 2 of read 100nA 2 of read 50nA ISET vs IMON High range 2 of read 1 A 2 of read 50nA 2 of read 5nA 2 of read 1 A 2 of read 500nA Low range 2 of read 100nA 2 of read 5nA...

Page 10: ...arts DT55xxE desktop power supply unit External AC DC power supply with separate power cord USB cable 10BASE T Ethernet cable Moreover an external PC is required supported OS Windows 7 or later Linux...

Page 11: ...r level application such as GECO2020 and the lower layer software libraries For more info please visit www caen it products firmware software section Power Supply Modules LabVIEW Instrument Driver Pow...

Page 12: ...k and set the virtual communication port associated to the module Select USB connection and the used port number set port as follows baud rate 9600 Data bits 8 Parity none stop bit 1 Flow control none...

Page 13: ...in the EEPROM and go back to Main Menu the DHCP server will automatically assign a new IP to the module at next Power On Firmware upgrade To upgrade the firmware download from www caen it unit page th...

Page 14: ...ded firmware Format EEPROM By typing F on Terminal Board Menu Fig 10 it is possible to access the format EEPROM menu Fig 12 Format EEPROM Menu Channels settings By typing D on Terminal Board Menu Fig...

Page 15: ...0x1000 At least one ch in Interlock CMD MON PAR MACADDR Board MAC Address CMD MON PAR IPADDR Board IP Address CMD MON PAR SUBMASK Subnet mask CMD MON PAR GATEWAY Gateway CMD MON PAR DHCPEN DHCP Enable...

Page 16: ...hannel in Temperature warning 80 C IS_OVT 0x400 Channel in OVT 125 C IS_KILL 0x800 Channel Killed IS_INTLCK 0x1000 Channel in Interlock CMD MON CH XX PAR RUP RAMP UP value CMD MON CH XX PAR RUPMIN RAM...

Page 17: ...nge of customers including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Ve...

Reviews: