GasAlertQuattro
User Manual
40
Confidence/Compliance Beep
• One beep every 1-120 seconds (beep
frequency is defined with
Note: (Confidence/compliance beep
automatically deactivates during a
low battery alarm, calibration fail,
bump test fail, self-test fail, and an
alarm condition.
Note: If
Low Alarm Acknowledge
is enabled, the audible alarm can be dis-
abled during a low alarm condition. The visual, audible, and vibrate alarms
remain active until the alarm condition changes or the detector deactivates.
Press
C
to acknowledge and deactivate the audible alarm. If the alarm esca-
lates to a high, TWA, or STEL alarm, the audible alarm reactivates.
If enabled, during an alarm condition
Latching Alarms
causes the low and high
gas alarms (audible, visual, and vibrator) to persist until the alarm is acknowl-
edged by pressing
C
and the gas concentration is below the low alarm setpoint.
The peak concentrations display continually until the alarm condition no longer
exists. Local regulations may require
Latching Alarms
be enabled.
Alarm
Screen
Alarm
Screen