Detector does not enter alarm.
Alarm setpoint(s) are set incorrectly.
. Define the alarm set-
points in Fleet Manager II.
Alarm setpoint(s) set to zero.
. Define the alarm set-
points in Fleet Manager II.
Detector is in calibration mode.
Complete the calibration procedure.
Detector intermittently enters alarm
without reason.
Ambient gas levels are near alarm setpoint
or the sensor is exposed to a puff of the
target gas.
Detector is operating normally. Use caution in suspect
areas. Check peak gas exposure reading.
Alarm setpoints are set incorrectly.
. Define the alarm set-
points in Fleet Manager II.
Sensor(s) requires calibration.
Refer to
.
Missing or faulty sensor(s).
Refer to
Features and options are not oper-
ating as expected.
Changes in Fleet Manager II.
Verify that the settings in Fleet Manager II are correct.
Table 14. Troubleshooting
Problem
Possible Cause
Solution