77
GasAlertQuattro
Troubleshooting
Calibration Troubleshooting
Table 16. Calibration Troubleshooting
Bump Test Troubleshooting
For bump test troubleshooting recommendations, refer to Table 15.
Error Screen
Problem
Solution
Error Screen
Problem
Solution
Detector deacti-
vates while trying
to calibrate
Detector does not
enter calibration.
Displays
OFF
and
then deactivates.
Activate the detector. When
in normal operation, initiate
calibration. Press and con-
tinue holding
C
(through a
series of screens) until
Calibration started
displays.
Cal IR Lock
Enabled
IR lock enabled
screen displays.
An IR device must be used to
calibrate (IR LInk or Micro-
Dock II station). For manual
calibration, refer to
automated
calibration, refer to the
Micro-
Dock II User Manual
.
Self-test Fail
If a sensor(s) fails
auto zero, an error
message displays
showing which sen-
sor(s) failed.
Replace the sensor or con-
tact
.
Calibration Error
Insufficient amount
of gas detected.
Verify the span gas values on
the cylinder match the span
gas values set for the detector.
Ensure gas is applied at a flow
rate of 250-500 ml/min.
Ensure cylinder is not empty
or past expiry. Replace imme-
diately if required. Replace the
regulator if required.