GasAlertQuattro
User Manual
56
The
Calibrate Device
dialog box displays.
Figure 15. Calibrate Device Dialog Box
9. Enter the span gas concentration values. The values entered
in the Calibrate Device popup must match the span concentra-
tion values on the gas cylinder, otherwise calibration will fail.
10. Click inside the checkbox for each sensor that will be
calibrated, and then click
. The following screen
displays.
step #8 in the
to
continue calibration.