l
Leave enough space (min. 1,5 m) around the vacuum system for
maintenance work.
l
If necessary anchor the vacuum system base frame to the floor
l
Reception and unpacking of the crates
l
Oil filling in the Roots MB vacuum pumps
l
Oil filling in the DP vacuum pumps
l
Glycol filling in the DP vacuum pumps
l
Set, fix and connect the Roots MB assembly (connection with the
inlet of the DS vacuum pump)
l
Install the piping bows on the DP/ MB vacuum pumps and the
troughs on the MB vacuum pump
l
Install the sensor wires and place the cables in the trough (power,
N2 solenoid valve, sensors) (connection with power supply)
l
Set and fix the H2O and N2 distributor (connection with the H2O
and N2 supply)
l
Cable and connect the MB Roots (power, sensors, H2O and N2)
l
Set and fix the DP pumps and connect the flexible piping to the
Roots MB (connection with the DS group)
l
Connect the pumps DP (power, N2 solenoid valve, sensors)
l
Connect the LCD control and switch the main contactor
l
Flow and H2O alarm adjustment, Danfoss
l
Start-up of the vacuum system (Start)
l
Flow and N2 alarm adjustment and final adjustment
WARNING
The vacuum system can be damaged by tensile or compressive
stresses on pipe connections!
l
If stresses could occur, use compensators to connect pipes!
WARNING
The vacuum system can be damaged by condensate!
l
Fit traps to inlet and discharge sided pipes to prevent condensate
getting into vacuum pumps!
Inlet connection
l
Make sure that the protection that was attached to prevent pene-
tration of particles during transport has been removed before the
vacuum system is connected to the vacuum line
CAUTION
Do not put hands into the inlet aperture.
Risk of body damage!
CAUTION
The intake of liquids or solid particles can lead to the destruction of
the vacuum system
If the sucked gas contains dusts or solid foreign bodies:
u
Make sure that a filter or protective grating is installed at the
extraction point
l
Make sure that the nominal diameter of the intake line is at least
equal to the diameter of the intake flange of the vacuum system to
prevent a drop in the performance of the vacuum system in the
case of a smaller cross-section
l
Make sure that the vacuum system is connected with leakproof
lines
CAUTION
When the intake lines have been connected, make sure that the sys-
tem does not leak. Leakages of dangerous substances must be pre-
vented!
l
Make sure that the intake lines are equipped with a shut-off device
upstream of the intake flange so that the flow of sucked gas can
be stopped
l
Make sure that the intake lines do not exercise any force on the in-
take flange. Mount bellows if necessary
l
The inlet flange has the following dimension:
–
DN 250 ISO-K for the Roots WY A075 inlet (DS 8163, 8264)
–
DN 100 ISO-K for the COBRA DP 600 C inlet (inlet backing
vacuum pump)
In the case of long intake lines the line cross-section should be larger
than the intake flange to prevent a drop in the performance of the
vacuum system. If you have any doubts, contact your Busch represen-
tative.
Discharge connection
CAUTION
Do not put hands into the outlet aperture.
Risk of body damage!
The following instructions for connection to the discharge only apply if
the sucked gas is discharged by the vacuum system into a suitable en-
vironment.
l
Make sure that the protection that was attached to prevent pene-
tration of particles during transport has been removed before the
vacuum system is connected to the vacuum line
l
Make sure that the nominal diameter of the exhaust line is at least
equal to the diameter of the exhaust flange of the vacuum system
to prevent a drop in the performance of the vacuum system in the
case of a smaller cross-section
Make sure that the vacuum system is connected with leakproof lines
CAUTION
When the discharge lines have been connected, make sure that the
vacuum system does not leak. Leakages of dangerous substances
must be prevented!
l
Make sure that the discharge lines are equipped with a shut-off
device upstream of the intake flange so that the flow of sucked gas
can be stopped
l
Make sure that no shut-off devices are mounted in the discharge
line
l
Make sure that the discharge lines do not exercise any force on the
intake flange. Mount bellows if necessary
l
The discharge flange has the following dimension:
–
DN 63 ISO-K for each backing vacuum pump DP 600
In the case of long discharge lines, the line cross-section should be
larger than the discharge flange to prevent a drop in the performance
of the vacuum system. If you have any doubts, contact your Busch
representative.
Cooling water connection
The cooling water is generally connected with a hose (leakproof
against oxygen).
l
Connect the cooling water hose to the vacuum system distributor
l
Connect the distributor to the cooling water supply
The cooling water outlet must be unpressurised.
COBRA DS 8163 A
Installation and start-up
0870772605
Page 17
Summary of Contents for COBRA DS 8163 A
Page 34: ......