TRANS CAL
Model 7280
Page 36
Given :
Load cell model 8524-6050
strain gauge 350
Ω
Range: 50
kN
Output
signal
: 1.5016
mV/V
Zero output without fitting parts:
0.0020 mV/V
We choose for the above sensor the menu „PAS nCON“ ---> passive sensor without control.
After pressing
ENTER
you get the following screen:
0 % LOAD
Or
after pressing
▼
you get the second entry menu (0 %)
NOM VAL
After pressing
ENTER
you can enter the zero value:
0.020 mV/V
After pressing
ENTER
you can enter the output signal (sensitivity (100 %) NOM VALUE, in this
example 1.502 mV/V. If the entering of the value ended the next is “save?”. Press
ENTER
to
complete the entry, the final value will then be stored in the 7280 measuring amplifier.
After
0 % LOAD
Now you can choose between two menus
100 % LOAD
or
the second entry menu (100 %)
NOM VAL
After pressing
ENTER
you can enter the 100 % value
1.502 mV/V
If the entering of the value ended the next is “save?”. Press
ENTER
to complete the entry, the
final value will then be stored in the 7280 measuring amplifier.
Load cell is connected and
the zero value is stored
with
ENTER
.
The characteristic data
is entered according
data sheet. In this case
0.020 mV/V.
Load cell is connected and
the zero value is stored
with
ENTER
.
Load cell is connected and
the 100 % value is stored
with
ENTER
.