TRANS CAL
Model 7280
Page 24
4.7 Adjustments in the LOGG mode
LOGG OUT
Here the data logger is switched off. the logg mode in the measur-
ing mode is switched to “LOGG OUT” as soon as the entire data
logger has been edited.
HAND
In this mode a measured value is written into the data logger when
the enter key was pressed at the 7280. By an additional logg of time,
there is a assignment for each measured value.
AUTO
In this mode in the adjusted delay a measured value is written into
the data logger. By the stored starting time there is a time assign-
ment for each measured value.
GRAPH
In this mode at a trigger event a measured value is written into the
logger. Since the trigger pulses occur in 10 ms raster, an additional
logg of the time is not possible. The flank of the trigger pulse must
stand on HIGH for 4 ms at last. Afterwards on LOW for at least 6 ms.
WINDOW
This mode reacts to in increasing and/or decreasing flanks. At an
increasing flank the logging of the measured values is started. From
now on the measured values are written in the data logger with 1
ms raster. A decreasing flank ends the recording.
Memory depth of data logger - see chapter 5.1
Trigger Signal
Read values