Model 7280
TRANS CAL
Page 33
Given :
Load cell model 8524-6050
strain gauge 350
Ω
Range: 50
kN
Output
signal
: 1.5016
mV/V
Zero output without fitting parts:
0.0020 mV/V
Adjustment procedure
The 7280 gets activated by pressing the
ENTER
key.
The following screen is displayed:
1 MEASUR
By pressing
▲
or
▼
you get to the next screen.
2 SYSTEM
Now press
ENTER
the following screen is displayed:
2.1 LANG
By pressing several times
▲
or
▼
until you get to the following screen:
2.5 PASS
By pressing
ENTER
you takes to the entering of the password:
0 0 0 0
After entering the password (9373), the sensor can be applied in the adjustment menu. (When
the beam is below a paragraph you can change this by pressing of
▲
or
▼).
You get to screen:
3 ADJUST
After pressing
ENTER
you get to the sensor selection 0 - 9, CURRENT, active, sensor test
SENSOR 1
7. Programmer Example