TRANS CAL
Model 7280
Page 18
2.7 LOGG
Data logger adjustments
LOGGMODE
Data logger configuration
LOGG off
Off-switch for logger operation, measuring mode possible, only.
Hand
With each keystroke on
ENTER
a log value is accepted.
AUTO
Adjustment of the measurement intervals for the automatic measured
value storing.
DIAGRAM
With each rising edge of the trigger signal, a log value is accepted.
WINDOW
The window operation is started by a rising trigger edge and can be
finished only by a falling edge. During this time, measured values
are stored in 1 ms raster in the data logger.
LOGGSEND
Log values are issued through the interface, press
ENTER
twice.
LOGGDEL
Log values are deleted after a safety query. Before deleting, please
assure that the data was received.
2.8 DATE
Data and time adjustment
DATE
Date
TIME
Time
3 ADJUST
Adjustmenu
This menu is active only, if the password has been entered in menu
mode 2.5 PASS. For the adjustment, the sensor must be connected
to the sensor socket.
Sensor__
Sensor number or the name of the sensor which shall be adjusted.
POINTDIS
Final measurement value of the sensor with decimal indication e.g.
1000; 100.0; 10.00; 1,000. The numerical values are variable, 4 digits
are available. Change to menu mode POINTSDIS. After pressing
ENTER
, the final measurement value of the sensor can be entered.
With keys
▲
and/or
▼
the first digit can be edited, go to the next
figure by pressing
ENTER
. When all 4 digits are edited, the decimal
point gets shifted by keys
▲
and/or
▼
. Press
ENTER
to complete
the entry, the final value will then be stored in the 7280 measuring
amplifier.
UNIT
Entry of the physical unit e.g. kg, Ncm, t, gr, kN, Nm, bar...
ADJUST
Sensor type selection (active with 100 % control signal, active without
100 % control signal, passive with 100 % control signal, passive
without 100 % control signal, 4-20 mA.
DESIG
Name of the sensor e.g. sensor 1, 2, 3, DR-2112, silo, tank, mixer,
scale 1, motor, test 1...