Model 7280
TRANS CAL
Page 19
4.2 Measuring with 7280
Measuring with 7280
selected sensor is:
Force sensor with meas. range
200 kN
Sens.:
2 mV/V
Designation e.g.:
Squeezer
Parameter set: Sensor no.:
3
Option:
100 % Cal. control
In order to be able to now measure with this sensor, the sensor „SQUEEZER“ is selected in
the measuring mode. During a measurement, a minimum and maximum value buffer can be
called up with the arrow keys, the in each case displayed minimum or maximum value can
be deleted with ENTER.
If the measured value should be sent through an interface to a pc or printer, the baud rate
must be adjusted to the receiver and the operating mode must be chosen in the case of SCI-
MODE. For example: HAND, a measured value with time is displayed during every keystroke
on ENTER while measuring.
Should data be logged, adjust operating mode in the logger mode, e.g. choose AUTO and
range time e.g. 10 sec and enter the measuring mode. A measured value with time is now stored
every 10 sec. In the mode LOGGSEND, these data then can be sent through the interface.
ENTER
4.2.1 Negative zero output
With no sensor parameter set, a negative zero output from the sensor can be entered (manu-
ally by hand). A negative zero output can only read in directly from the sensor with teach-in
(function adjustment).
4.2.2 Stored adjustment values
The stored adjustment value are not displayed. This will only stored inside the device and can´t
be changed manually.
Key assignment:
Key
Measuring mode
▲
Press = TARA
▼
MEASURING --- MIN --- MAX --- MEASURING
at MEASURING: send data
at LOGG:
store measured value
at MIN:
delete MIN
at MAX:
delete MAX
ESC
Back to menu / Press 3 sec. Press = Off