TRANS CAL
Model 7280
Page 20
4.3 Sensor parameters
After the password (9373) input in the system, the sensor can be applied in the calibrating menu.
Following parameters are available:
Sensor_
Sensor no 3
Sensor 0 - 9 possible
POINTDIS
200.0
Adjust meas. range and decimal point (max. 9999)
UNIT
_kN
1 – 3 digit unit possible
DESIGN
Squeezer
up to 8 digit name (or numbers) freely optional
ADJUST
PAS_nCON
Select passive without cal. control
0 % LOAD
Relieve sensor
0% value of sensor, indication 0 is assigned
100 % CON Autom. calibration 100 % value of sensor will be assigned to
indication 200 kN
SAVE
ENTER or ESC
Confirm or reject
4.4 ADJUST Adjustment menu
3 ADJUST
Adjustment menu
This menu is active only, if the password has been entered in menu
mode 2.5 PASS. For the adjustment, the sensor must be connected
to the sensor socket.
Sensor_
Sensor number or the name of the sensor which shall be adjusted.
POINTDIS
Final measurement value of the sensor with decimal indication e.g.
1000; 100,0; 10.00; 1.000; ...
The numerical values are variable, 4 digits are available. Change to
menu mode POINTDIS. After pressing
ENTER
, the final measure
-
ment value of the sensor can be entered. With keys
▲
and/or
▼
the
first digit can be edited, go to the next figure by pressing
ENTER
.
When all 4 digits are edited, the decimal point gets shifted by keys
▲
and/or
▼
. Press
ENTER
to complete the entry, the final value will
then be stored in the 7280 measuring amplifier.