Model 7280
TRANS CAL
Page 17
4. Manual Operation
4.1 Menu description
1 MEASURE
Measuring mode
SENSOR__
Sensor selection for measuring mode, sensor 0 - 9
Selection of the sensor parameter set of the measurement. The
sensor parameter set must comply with the connected sensor.
2 SYSTEM
All system parameters are stored in this menu column.
2.1 LANG
Language adjustment
The menu language is available in English, German, French and
Spanish.
2.2 INFO
Information query was factory-configured and is not modifiable.
SW no.
Firmware version
2.3 SCI
Interface configuration
BAUD
Baud rate adjustment; it must comply with the receiver (PC or printer).
SCI MODE
The interface configuration only refers to the measuring mode, not
the log mode!
SCI off
Interface switched off
HAND
When pressing
ENTER
in the measuring mode, a measured value
is displayed.
AUTO
1
Interval time, adjustable from 10 ms to 1 h (at high measuring rate
only possible with the highest baud rate).
TRIGGER
At rising edge on the trigger input, in measuring mode a measured
value is displayed.
INTERFAC
Interface selection
USB
USB interface is activated
RS232
RS232 interface is activated
2.4 RATE
Measuring rate
1
and average value
RATE
Measuring rate adjustment, selectable from 1/s to 1000/s. For fast
procedures (screw joint, insert press...) always select a fast measur-
ing rate, e.g. 1000/s, for very slow procedures select a small rate,
e.g. 1/s.
AVERAGE
Average determination, the numbers indicate the average determina-
tion by the amount of measurements. Applicable for e.g. vibrations,
control oscillations...
2.5 PASS
Password query
RATE
Entering password
9373
allows to change to menu 3
ADJUST
. Here,
the sensor parameters can be changed.
2.6 LCD
LCD contrast adjustment
If the LCD becomes unreadable by external influences, e.g. solar
radiation, heat or cold, the LCD contrast can be changed by press
-
ing key
▲
or
▼
.
1
If the SCI
→
AUTO interval time is shorter than the adjusted measuring rate, the same measured value
will be issued on the interface until a new value is entered.