Model 7280
TRANS CAL
Page 35
Given :
Load cell model 8524-6050
strain gauge 350
Ω
Range: 50
kN
Output
signal: 1.5016
mV/V
Zero output without fitting parts:
0.0020 mV/V
After pressing
ENTER
you get the following screen e.g.
_Nm
Now in this example the unit kN can be set by pressing
▲
or
▼
. If the beam is below the letter
this can be changed by pressing
▲
or
▼
. The changed letter is taken by press of
ENTER
.
Now return to the screen:
UNIT
By pressing 2 times
▲
you arrive the following screen:
DESIGN
After pressing
ENTER
you get the following screen:
SENSOR 1
Now it can be changed as for the designation of the UNIT e.g.
Place 1
or similar.
By pressing 1 times
▼
you arrive the following screen:
ADJUST
After pressing
ENTER
you get the following screen:
ACT nCON
ACT wCON
PAS nCON
PAS wCON
CURRENT
With
▲
or
▼
you can toggle through these screens.