burster
Digital Panelmeter
Model 9180
Page 2-21
2.4.3 Access to display settings
M
A
X
M
I
N
1
HOLD
TARA
2
TARE
RESET
ESC
LIMIT
MAX/MIN
DATA
ENTER
3
4
Fig. 2.4a
Press the "
ENTER
" button again. The supplementary display shows
20 CnFdSP
. Now you can
perform the corresponding configuration using the submenus.
You will find precise and detailled description of the
SCAL
and
TEACH
functions starting with
page 2-23.
Important:
Before each calibration delete to zero any existing TARA values.
To do this press TARE and RESET at the same time. First release
TARE and then release RESET !
2.4.4
Programming the display for the correct measurement
variables
Programming the display for the correct measurement variables
The display can be programmed from -99999 up to 99999 for process signals, strain gage
sensors (LoAD) and potentiometers. The thermometers do not need any programmng as the
input and measurement range are automatically set in accordance with the sensor type selected.
Here the settings can only be corrected with "Offset" (see page 2-10).
To program the measurement range you must enter 2 input values (
INPUT1
,
INPUT2
) and
2 output values (
DISPLY1
,
DISPLAY2
) in accordance with the two input values. The entry is
automatically completed by the decimal point setting.
Um in das Programm-Modul 20 zu gelangen, betätigen Sie
"ENTER"
, danach zweimal die
"
" -Taste.
Summary of Contents for 9180
Page 2: ...Model 9180 burster Digital Panelmeter Page ...
Page 3: ...burster Digital Panelmeter Model 9180 Page ...
Page 4: ...Model 9180 burster Digital Panelmeter Page ...
Page 8: ...Model 9180 burster Digital Panelmeter Page ...
Page 12: ...Model 9180 burster Digital Panelmeter Page ...
Page 104: ...Model 9180 burster Digital Panelmeter Page ...
Page 105: ...burster Digital Panelmeter Model 9180 Page 6 1 A P P E N D I X Example BASIC ...
Page 106: ...Model 9180 burster Digital Panelmeter Page 1 ...
Page 107: ...burster Digital Panelmeter Model 9180 Page 2 ...
Page 108: ...Model 9180 burster Digital Panelmeter Page 3 ...
Page 109: ...burster Digital Panelmeter Model 9180 Page 4 ...