Model 9180
burster
Digital Panelmeter
Page 2-8
Here are a few more instructions regarding the input configuration:
1. Process signals (Proc) and strain gage (DMS) sensors (LoAd)
When connecting process signals or strain gage (DMS) sensors the input signal range
must be specified.
Process signals:
There are 2 voltage ranges (0-1 V, 0-10 V) and 2 current ranges
(0 - 1 mA, 0 - 20 mA).
If you have selected "
VOLT
", 2 input ranges are offered.
Select between:
"
1V
"
for inputs 0 - 1,
"
10V
"
for process signals 0 - 10V.
The maximum voltage amounts to 10 V.
However, if you have selected
"AMP"
, you have the following ranges at
your disposal:
"
1mA
" for the input 0 - 1 mA,
"
20mA
" for the inputs 0 - 5 mA,
0 - 20 mA or 4 - 20 mA.
The maximum current amounts to 20 mA.
Strain gage sensors:
If the device is to be used as a DMS strain gage indicator, select one of
the following input ranges:
"
15 mV
", "
30 mV
", "
60 mV
","
300mV
".
Correct selection depends on the sensitivity of the DMS sensor. Sensors
with more than 300 mV input voltage may not be used.
Example: Characteristic value of the strain gage sensor (DMS) according
to the data sheet: 1.41 mV/V
Feed voltage (measured): 9.82 V
Signal from sensor computed at: U
(mV)
= 1.41
.
9.82 = 13.82
Selected input range is "15 mV".
Summary of Contents for 9180
Page 2: ...Model 9180 burster Digital Panelmeter Page ...
Page 3: ...burster Digital Panelmeter Model 9180 Page ...
Page 4: ...Model 9180 burster Digital Panelmeter Page ...
Page 8: ...Model 9180 burster Digital Panelmeter Page ...
Page 12: ...Model 9180 burster Digital Panelmeter Page ...
Page 104: ...Model 9180 burster Digital Panelmeter Page ...
Page 105: ...burster Digital Panelmeter Model 9180 Page 6 1 A P P E N D I X Example BASIC ...
Page 106: ...Model 9180 burster Digital Panelmeter Page 1 ...
Page 107: ...burster Digital Panelmeter Model 9180 Page 2 ...
Page 108: ...Model 9180 burster Digital Panelmeter Page 3 ...
Page 109: ...burster Digital Panelmeter Model 9180 Page 4 ...