Instruction Manual for Vertica-Homecare Bed
Page
69
Problem
Possible causes
Solution
Mains control
lamp in control
unit does not light
up
No power supply to socket
Mains cable damaged
Fuses in control unit de-
fective
Use a socket that works
properly
Replace the mains cable
Replace the control unit.
Inform the operator in order
to arrange for repairs
Drive runs for a
brief time only,
then stops
Drive overloaded
Structural obstructions in
the way of bed adjustment
Remove the overload in the
bed, retest
Remove obstructions; move
bed away from obstructions
(e.g. window sills, sloping
roofs)
Control unit partly
not functioning
One or more motors are
not connected/ electrical
plug connections are
disconnected
There is a serious problem
with the control unit.
For safety reasons, all
functions are locked.
Check electrical
connection and all
motors/plug-in connections;
Perform RESET; see Chapter
6.5.6; if it occurs again: Have
actuator system checked.
Inform the operator in order
to arrange for repairs
Height
adjustments and
tilting not possible
or only in one
direction;
acoustic signal
sounds during
adjustment
Control unit has
“forgotten” the actuator
positions
Perform RESET; see Chapter
6.5.6;
Summary of Contents for Vertica-Homecare
Page 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Page 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Page 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Page 83: ...Instruction Manual for Vertica Homecare Bed Page 84...