Troubleshooting
7.1
T
ROUBLESHOOTING
G
UIDE
The following table is a guide for rectifying faults:
Problem
Possible causes
Solution
Handset/
Actuator system
not functioning
(bed is
connected to
mains power
supply)
Mains cable not plugged
in
No power supply to socket
Plug not inserted properly
Actuators locked
Handset, mains cable or
control unit is defective
Insert mains cable; green
LED on the control unit must
light up
Check socket and fuse box
Check plug connections
Enable functions
Inform the operator in order
to arrange for repairs
Handset not
functioning,
adjustments are
not locked
Handset faulty
Control unit has detected
a fault and for safety
reasons has locked the
adjusting functions
Replace handset.
Perform RESET; see Chapter
6.5.6; if it occurs again:
Have actuator system
checked.
Inform the operator in order
to arrange for repairs
Battery-powered
operation not
possible
Battery not charged
There is no rechargeable
battery
Connect the bed to the
mains supply for approx. 8
hours
Constant alarm
sounds during
adjustment
Battery capacity depleted
Connect bed to the mains
supply to recharge battery
as soon as possible
Operation with
sufficiently
charged battery
only possible for a
short time
Limit of battery’s service
life reached
Replace battery.
Inform the operator in order
to arrange for repairs
Operation is not
possible despite
proper power
supply
Control unit has shut down
due to overheating
Control unit has detected
a fault and for safety
reasons has locked the
adjusting functions
Control unit defective
Observe max. duty
cycle: intermittent duty 2
min ON/18 min OFF; replace
the control unit.
Perform RESET; see Chapter
6.5.6;
Replace the control unit.
Inform the operator in order
to arrange for repairs
Manual
emergency
lowering of
backrest is not
possible
Bowden cable is too loose
or not secured
Bowden cable is kinked
Readjust at the release lever
or secure
Install new Bowden
cable. Inform your operator
about any necessary repairs
Summary of Contents for Vertica-Homecare
Page 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Page 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Page 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Page 83: ...Instruction Manual for Vertica Homecare Bed Page 84...