![burmeier Vertica-Homecare Instruction Manual Download Page 57](http://html1.mh-extra.com/html/burmeier/vertica-homecare/vertica-homecare_instruction-manual_2832796057.webp)
Instruction Manual for Vertica-Homecare Bed
Page
58
Inspection Report following an Inspection of Electromedical Equipment According to
Din EN 62353 (VDE 0751-1): 2015-10 - Page 1 of 3
Customer / med. facility / practice:
Address:
Carried out:
Repeat inspection
Inspection prior to initial operation(reference value)
Inspection following repairs/servicing
Device type:
care bed
hospital bed
Protection class:
I
II
Bed type:
Vertica-Homecare
Inventory number:
Location:
Serial number:
Manufacturer
Joh. Stiegelmeyer GmbH & Co. KG
User-specific parts:
None
Testing equipment used (type/inventory no.):
1.
MP-RL 93/42 Classification:
Class I, Type B
2.
I.
Visual inspection
WHAT?
HOW?
ok
Not
ok
Description
of defect
Visual inspection of the electrical components (if applicable)
Stickers and type plates
Present, legible
Control unit/transformer housing Correct position, damage
Motor housing and lifting tubes Correct position, damage
Handset/foot pedal
No damage
Motor cable, handset cable,
mains cable
Damage, routing
Mains cable holder
Fixed position
Plug and plug cover on control
unit
Available, correct position
Visual inspection of the mechanical components (if applicable)
Stickers and type plates
Present, legible
Patient lifting pole, location
sleeves
Grab handle with strap
Without damage, deformation, or
cracks
Chassis
Damage, deformation
Castors
No damage
Mattress base
No damage
Welded seams
Split welded seams
Gas springs, dampers,
hydraulics
No damage, sealing
Safety sides
Without damage, crack
deformation; authorisation for this
bed
Connecting elements (screws,
bolts, nuts, safety caps)
Fixed position, completeness
Wearing parts, joints
No damage, severe wear
II.
Electrical measurement according to DIN EN 62353 (VDE 0751-1): 2008-08:
Threshold
Measured
value
Leakage current of the device
, DC/Diff.
(place bed with conductive castors in a way that it
is insulated).
1.
Plug the bed mains cable in the test socket on
the measuring instrument.
2.
Connect sensor of measurement device to a
blank, conductive part of the mattress base
(screw).
3.
For the duration of the measurements, activate
the motors using the handset
0.1mA
mA
X
X
Summary of Contents for Vertica-Homecare
Page 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Page 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Page 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Page 83: ...Instruction Manual for Vertica Homecare Bed Page 84...