Inspection Report following an Inspection of Electromedical Equipment
according to DIN EN 62353 (VDE 0751-1): 2015-10 - Page 2 of 2
III.
Performance inspection
WHAT?
HOW?
ok Not
ok
Description of
defect
Performance inspection of the electrical components (if applicable)
Battery powered; capacity of
battery (optional equipment)
Requirements: Battery is charged
+ bed is disconnected from
power supply: Test: Load bed with
approx. 80 kg (=1 person); min. 2
cycles height adjustment
Up/Down until cut-out must be
possible
Emergency operation using
batteries; battery capacity; to
check: connect bed to mains
supply for 24 h, then run test.
Minimum of 3 height adjustment
cycles with a load of approx. 80
kg on the bed.
End of travel cut-out for motors Automatic cut-out
Handset, foot pedal;
operational controls, locking
functions
Test according to instruction
manual
No "rattling" when shaken
Motors
Abnormal noise level Emergency
release backrest motor
Control unit/transformer and
motors
Test according to instruction
manual
Strain relief of mains cable
Mains cable firmly fastened
Performance inspection of the mechanical components (if applicable)
Joints and pivots
Smooth operation
Grab handle with strap
Secure position
Castors
Brakes, securely engaged brake
Safety sides
Locking in place, release
Accessories (e.g. patient lifting
pole, grab handle, external
safety sides)
Fixing, damage, suitability
Overall inspection result:
Defects/remarks:
No safety or functional defects were detected
No direct risk, the defects detected can be rectified quickly
Appliance must be taken out of circulation until the defects have been rectified!
Appliance does not conform to requirements – modification/replacement of
components/decommissioning recommended.
Test approval sticker applied:
yes
no
Next inspection date:
Documents that form part of this inspection report:
Checked
Date:
Name:
Signature:
Test approval
sticker
applied:
Date:
Name:
Signature:
Address/stamp of responsible company
Summary of Contents for Vertica-Homecare
Page 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Page 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Page 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Page 83: ...Instruction Manual for Vertica Homecare Bed Page 84...