Page 33
CALIBRATE FLOW (cont.)
Procedures to calibrate the dilution flow rate:
1. Place a container, accurately graduated in ounces,
and with a minimum capacity of 1800 milliliters,
below the funnel on the brewer.
2. Press and hold the ® pad until the display reads
WARMER OFF
. Release the ® pad. Press and
release the ® pad until the display reads
CALI-
BRATE FLOW
.
3. Press and release
YES
to advance to the
SPRAY-
HEAD CAL?
function screen. (Pressing
NO
in the
CALIBRATE FLOW
screen will advance to the next
function screen,
BREW COUNTERS).
4. Press and release
NO
to advance to the
DILUTION
CAL?
screen.
5. Press and release the
YES
pad. The display should
read
CONTAINER READY?
If the container is un-
der the funnel, press
YES.
The display should read
CALIBRATE DILUTE ... PRESS BREW TO START.
6. Press and release the "BREW" pad. The display
should read
CALIBRATE DILUTE ... 60 SEC TO
FINISH.
The 60 second timer on the display will
count down to zero. When the counter reaches
zero, the display will change to
ENTER MI.,
along
with a number measured in milliliters per minute.
7. Measure the amount of water in the container, and
using
(-)
or
(+)
, adjust the amount on the display
to match the amount in the container. Then press
DONE
.
8. The display should now read
NEW DILUTE FLOW
,
along with the correct flow rate of the dilution in
milliliters per minute. After about 5 seconds, the
display will return to the
CALIBRATE FLOW
screen.
9. To exit the
CALIBRATE FLOW
function, press and
release
NO
to advance to the next function screen,
or press the "ON/OFF/LOWER" pad located on the
front switch panel to exit the
CALIBRATE FLOW
function and return to the
MAIN SCREEN.
ENTER MI XX.X
(-) DONE (+)
NEW DELUTE FLOW
XX.X MI/M
EXIT TO THE
NEXT FUNCTION
SCREEN
COUNTS DOWN
TO ZERO
BREW PRESSED
AFTER 1 MINUTE
CALIBRATE FLOW?
NO YES
CONTAINER RDY?
NO YES
CALIBRATE DILUTE
PRESS BREW START
CALIBRATE DILUTE
60 SEC TO FINISH
DILUTION CAL?
NO YES
SPRAYHEAD CAL?
NO YES
P3224
PROGRAMMING FUNCTIONS - LEVEL 2 (cont.)
37467 110904