3. Characteristics
3.1. Measuring characteristics
at alcohol content measurement:
Range (% by volume )
0 -15
15 - 25
Resolution
0,01%
0,01%
Repeatability
±0,05%
±0,1%
Measurement accuracy (absolute)
±0,1%
±0,2%
at boiling temperature measurement:
Range
0 -120 °C
Resolution
0,001 °C
Repeatability
±0,02 °C
Measurement accuracy (absolute)
±0,03 °C
at atmospheric pressure measurement:
Range
700 - 1100 hPa
Resolution
0,1 hPa
Repeatability
±0,2 hPa
Measurement accuracy (absolute)
±0,3 hPa
Quantity of sample measured (cleaning and analysis) - about 150 ml
Time per measurement < 6 min.
3.2. Operating conditions
Ambient temperature........................................................................... 15-35 °C
Relative ambient humidity.................................................................. 30-80 %
Temperature of sample measured........................................................ 10-45 °C
3.3. Power source
Stationary......................................................... 230/110 V AC +10/-15% 50/60 Hz
Power consumption ......................................... < 200VA
3.4. Additional characteristics
Overall dimension (H x W x D) ...................... 485 х 230 х 220 mm
Weight .............................................................. 6 kg
4. Installing and preparation for work
Place the meter in vertical position on a smooth and level surface (table, writing desk, etc.).
Connect the cooling water to water supply network or another source of cooling water as shown in
Fig. 3.
The source of cooling water shall meet the following requirements:
- Temperature - not higher than 25°C
- Flow rate - from 0.5 to 1.0 L/min.
Put both Vinyl hoses 8mm x 1.5 m and tighten the fixing clamps at the inlet and outlet of the
cooling system.
Connect the cooling system inlet to a water supply network or another source of cooling water as
shown in Fig. 3. Consult a technician for supplying of suitable part.