210705AS.CSV
where
21 -Year
07 - Month
05- Date
A - Hour (encoded) - in this case 10
S - Minute(encoded) - in this case 56
Notes:
1).The creation time of the data collection files on the external USB flash drive corresponds to the
time of their uploading (copying) .
2). Data collection files stored in the SD card built into the device are kept up to 90 days.
3). Obsolete data collection files (older than 90 days) saved in the built-in SD card are deleted from
the Startup procedure (see item 5.1.)
This mode allows the data collection files to be uploaded (copied) to the external USB flash-pen
drive, which must be plugged into one of the USB connectors located on the back panel of the
device.
To select the "Data collection upload" mode from the "File management" menu (see item 5.4.7.), select the
"Data collection upload" menu option using the <Down arrow> / <Up arrow> buttons and then confirm with
the < OK>.
A panel will appear on the display in which the user have enter the period of time (start and end date, month
and year) for which he wants to upload the corresponding files.
In this panel, use the <Left arrow> / <Right arrow> buttons to select the field for the start or end date, month
and year.
Once you have selected the field you want to change, use the <Down Arrow> / <Up Arrow> buttons to
change the value.
After making all the changes, press the <OK> button to start the uploading (copying) files process.
The files will be copied to the USB flash pen drive in the DATFILES folder.
Example:
K:\EBULL\U0200001\DATFILES
where :
К
- Device label (USB flash pen drive )
EBULL - Ebulliometer folder
U0200001
- Ebulliometer serial number ( see item 5.4.9. )
DATFILES
- Data collection files folder
During the copy(upload) process is in progress, the following message will be shown: