From this screen you may select the icon "Measurement" or "Settings" using the buttons <Left
arrow> / <Right arrow>.
5.3. "Measurement" mode
"Measurement" mode is designed to determine the boiling point of the sample(wine) and calculating the
volume percentage of ethyl alcohol.
Attention: During this procedure tap water must flow through the cooling system ( see point 4).
Before starting "Measurement" mode the measuring system should be flush with wine to be measured.
Close the discharge valve (see Fig. 2), and fill the measuring system to the top with wine to be measured
(about 100 ml). Then open the discharge valve to evacuate the measuring system. This way the measuring
system will be flushed and prepared for measurement.
Close the discharge valve, measure another 50 mL from the sample and pour it into the inlet of the
measuring system.
Adjust the wine volume to the printed mark ( see Fig. 6 ) on the reactor by opening the valve to drain the
excess wine. During the volume adjustment, the mark should be at your eye level.
Fig. 6