Once "Edit the note" menu option is selected (see item 5.4.2.) on the screen will appear the text for the
selected wine type, saved in the non-volatile memory in the device.
The blinking cursor indicates the position at which the next character will be entered. Use the <Left Arrow> /
<Right Arrow> buttons to change the cursor position. To fill in the note text, use the symbols "a" ... "z", "A"
... "Z", нумберс "0...9" and the punctuation marks as well.
To confirm the entered text press the button <OK>. This way the text will be saved in the non-volatile
memory in the device.
If you do not want the changes to be saved, press the <MODE> or <HOME> button.
5.4.3. "Cleaning" mode
A regular and proper cleaning of the meter measuring system will assure accurate and reliable measurement
results. You may apply different cleaning solutions and procedures depending on the samples tested:
- In case you are measuring dry wines only it is enough if at the end of the working day you run the device
few times (2-3) in "Cleaning" mode, using distilled water as a cleaning solution.
- In case you are measuring sweet wines the cleaning procedure is to be made at the end of the working day
by running the device few times (2-3) in "Cleaning" mode, using 5% of NaOH solution as a cleaning
solution.
Distilled water should be used then to flush the measuring system in order to remove NaOH remains.
Before starting the "Cleaning" mode, close the drain valve (see Fig. 2), measure out 50 mL of cleaning
solution (distilled water or 5% of NaOH solution) and pour it into the input of measuring system. The
volume must be up to the level of the printed mark placed on the reactor. Drain excess through the drain
valve. During this procedure, the reactor printed mark must be at eye level (see Fig. 6).
To start the "Water" mode form the "Settings" menu select "Cleaning" option, using the buttons <Up arrow>
/<Down arrow> and confirm with <OK> button.