Reading / Setting the Configuration Register
Note: Not available in mode 0 and 0 +128
Syntax:
K <RNo.> [ = <Value> ]
K
Command
<RNo.>
Relay number 0 or 1
= <Value>
Set optional configuration register to value (0 -- 255), otherwise read
The response of each relay can set via the configuration register. The setting has also an ef-
fect on the limit monitoring. Only when one relay is set to tolerance range (value >5), will
the second limit value also be fetched for processing.
Following options can be selected:
Value
Relay Response
0
passive, responds to no limit value, only to R command
1
relay always active when instrument is on
2
activates when limit value 1 is reached
3
activates when limit value 2 is reached
4
activates when limit value 1 is stepped under
5
activates when limit value 2 is stepped under
6
active as long as the measured value is between limit value pair 1
7
active as long as the measured value is between limit value pair 2
8
active when measured value is outside limit value pair 1
9
active when measured value is outside limit value pair 2
Example: Relay 0 shall respond only to interface command.
Computer:
K0=0
Instrument: OK
Read configuration relay 0
Computer:
K0
Instrument: 0
Note: This value can also be set directly in menu
Func.
at
rEL.1
/
rEL.2
.
Manual PM984