Reading / Setting the Measured Value
Note: Not available in the mode 0 and 0 +128
Syntax:
W [L | H | M ]<CNo.> [ = R | <Value>]
W
Command
L
Minimum value, optional, otherwise current measured value
H
Maximum value, optional, otherwise current measured value
M
Average value, optional, otherwise current measured value
<CNo.>
Channel number 0
= R | <Value>
Set to optional value (floating comma number), otherwise read
The instrument provides the following measured values:
Current measured value
Minimum measured value
Maximum measured value
Average measured value (max. over 3,3 million values)
The values are replied in ASCII format with sign followed by decimal point. In addition,
the possibility to set and reset the values is given in order to restart or re-initialize the gen-
eration of maximum value, minimum value and average value. The values must be given
as integer value (range of 32767 to -32768). An equal number of decimal posi-
tions are assumed like the ones that were set for the scaling.
Example: Read current measured value
Computer:
W0
Instrument:
+5788 mm
Reset average measured value
Computer:
WM0=R
Instrument:
OK
Read average measured value
Computer:
WM0
Instrument:
+3762 m/s
Manual PM984
State July 2002 - All subject to change without notice
17