C-12
Model 0254
Section C Serial Communications Protocol
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
C-4-2-2 Send Channel Input Port Programmed Values
This command enables a terminal operator or host to acquire the present
state of input port programmed values. Valid input port numbers (xx) are
[1, 3, 5, 7].
HOST SEND
AZ[.xx]V<cr>
Non-network
AZ[yyyyy.xx]V<cr>
Network
RESPONSE
PROGRAM VALUES - Channel 1 - Port 01
<04> Measure Units ml
<10> Time Base min
<03> Decimal Point x.xx
<27> Gas Factor 1.000
<28> Log Type Off
<00> PV Signal Type 0-20mA
<09> PV Full Scale 20.00 ml/m
Each of the input port programmed operating values can be individually
queried or changed in accordance with the procedures in Section C-5-1
and C-5-2.