2-3
Model 0254
Section 2 Installation
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
2-9 Cable Requirements
For compliance with the EMC directive 89/336/EEC, the equipment has to
be installed with shielded signal cables which are overall completely
screened with a shield of at least 80%. Sub-D connectors used must be
shielded with a metal shield. The cable screen should be connected to the
metal shell and shielded at both ends over 360°. The shield should be
terminated to earth ground.
The optional power supplies available from Brooks are always supplied
with a power cord that meets all agency certifications and has a protective
conductor for grounding purposes. Any replacement cords must have
similar construction and be certified by a recognized nartional test
laboratory.
2-10 Installation Instructions
Do not use cleaning agents other than water because this might affect
color and marking of the equipment.
2-10-1 Panel Mount Installation Instructions
1. Cut a hole according to the cut-out dimensions shown in Figure 2-1 on
p. 2-4.
2. Pass the Model 0254 enclosure through the cut-out.
3. Position one of the two brackets included in the Panel Mount kit on the
side of the enclosure, while aligning the two holes at the top and bottom
of the bracket flange with the holes at the corners of the back of the
enclosure. Secure the bracket to the enclosure with two of the screws
provided in the kit.
Do not over-tighten these attachment screws.
CAUTION
!
4. Secure the second bracket to the other side of the enclosure.
5. Tighten the panel mounting screws in the middle of both brackets to
secure the enclosure to the panel.
Do not over-tighten these attachment screws.
CAUTION
!
6. Connect the power supply cable with the D-connector to the power
connector on the back of the enclosure.
7. Connect the power supply AC cord to a power outlet.
8. Connect the cables to the RS-232 and Channel connectors, as
approprate for the application.