3-6
To mount the transmitter:
1. Identify the components shown in Figure 3-3. For dimensions, see Appendix A.
2. If desired, re-orient the transmitter on the bracket.
a.
Remove the junction end-cap from the junction housing.
b.
Loosen each of the four cap screws (4 mm) inside the junction housing.
c.
Rotate the bracket so that the transmitter is oriented as desired.
d.
Tighten the cap screws, torquing to 30 to 38 in-lbs (3 to 4 N-m).
e.
Replace the junction end-cap.
3. Attach the mounting bracket to an instrument pole or wall. For pipe mount, two user-supplied
U-bolts are required. Contact Brooks to obtain a pipe-mount installation kit if required.
Figure 3-3 Transmitter Components – Field-mount Transmitters
3.4 Grounding the Transmitter
Improper grounding could cause measurement error.
To reduce the risk of measurement error:
•
Ground the transmitter to earth, or follow ground
network requirements for the facility.
•
For hazardous area installations in Europe, refer to
standard EN 60079-14 if national standards do not
apply.
Ground screw
Mounting bracket
Main enclosure
Junction housing
Mating connector
socket
Mating connector
Junction end-cap
4 X Cap screws
(4 mm)
Conduit opening
for 4-wire cable
CAUTION
Instruction and Operation Manual
X-CM-QmBS-eng
Part Number: 541B052AHG
June, 2008
Section 3 Installing Field-Mount Transmitter
QmBS Sensor / QTA Transmitters