6-3
Figure 6-3 RS-485 Point-to-Point Wiring
Figure 6-4 HART Multidrop Wiring with SMART FAMILY
™
Transmitters and a Configuration Tool
Note: The RS-485 communication wires must be shielded.
Other devices
RS-485/A
RS-485/B
Primary
controller
Multiplexer
HART-compatible
host or controller
HART-compatible
transmitters
SMART FAMILY
™
transmitters
Note: For optimum HART communication, make sure the
output loop is single-point-grounded to an instrument-grade
ground.
24 VDC loop power
supply required for
passive transmitters
600
Ω
maximum resistance
250
Ω
minimum resistance
QTA FM AN
transmitter
Instruction and Operation Manual
X-CM-QmBS-eng
Part Number: 541B052AHG
June, 2008
Section 6 Wiring FM AN Transmitter
QmBS Sensor / QTA Transmitters