2.1 TOUCHSCREEN
CONT.
2.1.8 SETTINGS
(USER ACCOUNT)
2. Operation Instructions
Cont.
Settings
Exit
UV Decontamination
Defrost Control
Single
Dual
Compressor
Quick Freeze
-45.0
°C
Chamber
21.0
°C
Specimen
-36.5
°C
End Session
Settings
00002
<< MICROTOME >>
Rewind
Advance
HandWheel Counter
Light Off
Settings Button on
Home Screen comes
to this screen
UV Decontamination
duration settings
Defrost duration and
time settings
Return to
Home Screen
HIGH SPEC ONLY
Select machine to run
in either single or dual
compressor mode.
When running in Single mode the cryostat will
automatically switch between one compressor
and the other in order to reduce wear on the
compressors. Please note that there will be a pause
of 40 to 70 seconds between the compressors
starting in order to reduce current surges.
In the Dual mode, full advantage can be
taken of the extra refrigeration power
available, with faster cool down times
and an extra 5ºC can be achieved in the
cryostats freezing performance.