Figure 17 – Password Change
To disable the password, proceed as indicated above and select 0000 as the new
password.
To return to the previous menu, select the position “Previous Menu” and press the
knob.
8.5
Calibration
Menu
The calibration menu allows the calibration of the pH sensor.
8.5.1
Importance of Calibrating
The SANUS equipment is delivered with an electrochemical sensor to do the pH
readings. The sensor calibration is meant to translate the readings performed by the
sensor to the system, so that it acts correctly and responds promptly only and every
time it is necessary.
The SANUS software has been designed to assure an efficient calibration and this is the
reason why it is a slow procedure!
8.5.2
When to Calibrate
The equipment is precalibrated in factory. Nonetheless, it is recommended to check
the sensor readings upon installation and, if necessary, run the calibration procedure
once more. Every 6 months, or, exceptionally, when anomalous readings occur, the
procedure should be repeated.
8.5.3
Who should perform the calibration
The calibration menu should only be accessed by technically skilled personnel with the
proper knowledge of the calibrations procedures of the various tpes of sensors.
8.5.4
How to calibrate
Figure 18 – Calibration Menu
Before initiating make sure you have every tool and material necessary at hand.