14
Doc. HD40-0007-02 En-Eu L-4
3. The
Z1
will power up and cycle through a series of
startup displays.
4. After about 10 seconds, the
Z1
will enter standby
mode. The display will read 0.0. (Figure 8)
NOTE:
Remove the
Battery
from the
PowerShell
when the
Z1
is not in use, unless
the
Power Supply
is plugged into the
PowerShell
to charge the
Battery
. Otherwise
the
Z1
and
PowerShell
will drain the battery completely within 2-3 days.
Using Your Z1
LCD Display
Treatment Pressure Display
The current treatment pressure is shown on the LCD display. Once set, the prescribed treatment
pressure will be retained in the
Z1
’s memory even during loss of power. For example, a treatment
pressure setting of 10.0 cmH
2
O is displayed in Figure 9.
Ramp Progress & Ramp Time Remaining Indicators
Ramp Mode allows you to gradually acclimate to the prescribed treatment pressure. Ramp Mode
gradually increases the pressure from a low initial starting pressure to the prescribed treatment
pressure over a programmed interval of time. See “Programming Ramp Mode”.
If you program the
Z1
for Ramp Mode, the Ramp Progress and Ramp Time Remaining indicators
will appear on the LCD display. For example, Figure 9 indicates that the Ramp is just over half
complete, with a Ramp Time remaining of 10 minutes.
Figure 8
Treatment Pressure
Display
Ramp Progress
Ramp Time
Remaining
AC Power Indicator
Battery Status Display (PowerShell Only)
General Status Display
Fault
Reminder
Figure 9