WARNINGS:
• Do not connect or insert any device, cable or articles into the Micro USB or Micro
SD ports of your
Z1
except as described in this User Guide. Connecting other
devices could result in injury to you or damage to the
Z1
and void the warranty.
• Do not locate your
Z1
in an unapproved place or where the air intake can be
blocked during treatment. This will lead to overheating of the
Z1
. If a blocked intake
should occur during operation, the unit may stop supplying pressurized air and
cause a fault condition. Refer to “Troubleshooting” in this User Guide to interpret
this fault and instructions for corrective action.
• The
Z1
is designed to deliver pressures from 4 to 20 cmH
2
O. In the unlikely event of
certain fault conditions, pressures of up to 30 cmH
2
O are possible.
• Blocking the
CPAP Tube
while in operation could lead to overheating of the
Z1
.
• In the clinical environment, any personal computer that is used with your CPAP
system must be at least 1.5 meters (5 ft) away from, or at least 2.5 meters (8 ft)
above, the patient. It must also comply with the relevant test standard. For personal
computers the international standard is IEC 60950 or equivalent.
CAUTIONS:
• Be careful not to place your
Z1
where it can be bumped or where someone is likely
to trip over the power cord.
• Make sure your
Z1
,
Power Supply
,
PowerShell
and
Battery
are kept clean, dry, and
free of dust.
• Always make sure the power cord and plugs are in good condition and the
equipment is not damaged before use.
• Make sure the area around your
Z1
and accessories is dry and clean.
• During partial (below rated minimum voltage) or total power failure, therapy
pressures will not be delivered. When power is restored, your
Z1
will enter standby
mode and all prior settings will be retained in memory.
• Follow the storage and cleaning instructions provided with your
CPAP Tube
and
mask.
• The air filter should be changed at intervals specified in this User Guide.
Accumulation of filtered particles over time can reduce the efficiency and shorten
the life of any CPAP device, including the
Z1
. Replacement filters are available at
www.breas.com.
• Do not wash the air filter. The air filter is not washable or reusable.
• The
End Cap
protects the
Z1
in the event of a minor collision. Always ensure that
the
End Cap
is securely fitted in accordance with instructions at all times. The
Z1
may not operate if the
End Cap
is not properly installed.