18
Doc. HD40-0007-02 En-Eu L-4
Z-Breathe™ Algorithm Selection
Z-Breathe is a proprietary algorithm engineered in the
Z1
to smooth out the breathing cycle while
enjoying the benefits of CPAP therapy. Typically users experience pressure swings while breathing
on a fixed pressure CPAP device. Pressure swings occur as a spike in pressure during exhalation
and a drop in pressure during inhalation. Z-Breathe is designed to eliminate the spikes and drops
while keeping the pressure constant, thereby delivering more comfortable therapy.
Pressure swings can vary between users and this is due to the lung capacity of the person
breathing on the machine. Larger individuals will typically have greater lung capacity that will
require a more aggressive setting and for this reason we have created 3 levels of Z-Breathe:
—
CPAP 1
offers mild pressure swing relief
—
CPAP 2
is the
Default Setting
and offers moderate pressure swing relief
—
CPAP 3
is the most aggressive setting for pressure swing relief
You can change the setting of the Z-Breathe Algorithm when your
Z1
unit is in Standby Mode
(pressure of “0.0” indicated on the display). To change the setting, hold down both the
Up
and
Down Arrow Buttons
at the same time until you see “CPAP – 1”, “CPAP – 2” or “CPAP – 3” flashing
on the display. Press the
Up
or
Down Arrow Buttons
until you observe your desired setting on the
display. To lock-in the setting, hold down the
Start/Stop Button
until the “CPAP – X” indicator no
longer flashes on the display.
The Z-Breathe settings in relation to the fixed pressure are illustrated in the graph below. We
encourage you to experiment with the settings to achieve maximum comfort.
Pressure
Swi
ng
10 cmH
2
O
Z-Breathe 1
Z-Breathe 2
Z-Breathe 3