Troubleshooting
Problem
Possible Causes & Solutions
No Display
Power is not present. Ensure the
Power Supply
is
connected and the power outlet has power.
Loose connection of the
Power Supply
at the
Z1
or loose connection of the
Power Cord
at the AC
Adapter. Check the connections.
Z1
is not locked into position in the
PowerShell
.
Slide the
Z1
into the
PowerShell
until it clicks and
locks into place.
Battery
is not locked into position in the
PowerShell
. Slide the
Battery
into the
PowerShell
until the
Battery Shuttle
clicks and locks into place.
Battery
is not charged. Charge the
Battery
.
A power supply not designed for the
Z1
is con-
nected to the
Z1
. Remove the power supply unit
and replace with a Breas approved unit.
Treatment Will Not Start
Check the display for a fault condition. Refer
to the Fault Codes table in “Fault Conditions”.
Contact Breas or your health care provider.
Insufficient Air Flow Delivered from
Z1
Air filter is dirty or clogged. Replace air filter.
Air intake is blocked. Clear blockage.
CPAP Tube
is not connected properly. Check
CPAP
Tube, Tube Adapter
and connections to your
Z1
and your mask.
Ramp Mode is in use. Wait for air pressure to reach
prescribed treatment pressure or stop treatment
and adjust Ramp time.
CPAP Tube
is blocked, pinched or punctured.
Unblock
CPAP Tube.
Check the
CPAP Tube
for
punctures. Replace a damaged
CPAP Tube
imme-
diately.
Mask and/or headgear are not positioned correct-
ly and mask is leaking. Adjust position of the mask
and headgear until leak ceases.
Incorrect
CPAP Tube
selected. Do not use tubing
which was not recommended by your clinician.
Check the display for a fault condition. Refer to
fault codes in “Fault Conditions”.