74 Alarms
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
Low Leakage Alarm
I
TEM
D
ESCRIPTION
Definition
A low leakage alarm will be given when the meas-
ured flow is lower than the expected leakage flow
at set pressure.
Possible cause
• Incorrect patient circuit leakage.
• Obstructed or occluded patient circuit.
Setting
Self adjusting
Ventilator action
The Vivo 40 tries to continue delivering breaths
according to settings.
Indication
The alarm is given audibly with a tone and visibly
by the yellow alarm LED and a display message.